Abstract
In this study, an ultrasensitive biosensor based on a novel electrochemiluminescence nanosurface energy transfer (ECL-NSET) system was developed for procalcitonin (PCT) detection. N-(aminobutyl)-N(ethylisoluminol) (ABEI) as ECL donor was externally cross-linked with ferritin (ABEI-Ft), whose ECL emission was boosted by ferritin nanocore in the presence of hydrogen peroxide. Then, single gold nanoparticle (Au NP) as ECL acceptor was in situ reduced on ABEI-Ft surface to fabricate a donor-acceptor nanostructure (ABEI-Ft@Au). ECL quenching occurred in ABEI-Ft@Au correlates well with the NSET theory rather than FÖrster resonance energy transfer (FRET). To improve sensitivity of biosensor, HWRGWVC heptapeptide (HWR) was utilized to capture antibody Fc portion via specific interaction to realize site-oriented immobilization. After connecting with ABEI-Ft@Au via Au-S bond, HWR improved the incubation efficiency of antibody with a better maintained biological activity. Under optimal conditions, the proposed biosensor provided a quantitative readout to PCT concentration in the range of 100 fg/mL- 50 ng/mL with a detection limit of 41 fg/mL. With favorable specificity, stability and reproducibility, this highly efficient ECL-NSET system will inspire more interests in nanometal-based ECL quenching mechanism studies and developments of new methods for site-oriented antibody immobilization improve the sensitivity.
Funding Information
  • National Natural Science Foundation of China (21505051, 21575050, 21627809, 21427808, 21777056)

This publication has 41 references indexed in Scilit: