Single‐shot halfk‐space high‐resolution gradient‐recalled EPI for fMRI at 3 tesla

Authors:

Abstract
Halfk‐space gradient‐recalled echo‐planar imaging (GR‐EPI) is discussed in detail.T2* decay during fullk‐space GR‐EPI gives rise to unequal weighting of the lines ofk‐space, loss of signal intensity at the center ofk‐space, and a point‐spread function that limits resolution. In addition, the long readout time for high‐resolution fullk‐space acquisition gives rise to severe susceptibility effects. These problems are substan tially reduced by acquiring only half ofk‐space and filling the empty side by Hermitian conjugate formation. Details of the pulse sequence and image reconstruction are presented. The point‐spread function is 3½times narrower for half than fullk‐space acquisition. Experiments as well as theoretical con siderations were carried out in a context of fMRI using a whole‐brain local gradient and an RF coil at 3 Tesla. Using a bandwidth of ±83 kHz, well‐resolved single‐shot images of the human brain, as well as good quality fMRl data sets were obtained with a matrix of 192 × 192 over 16 × 16 cm2FOV using halfk‐space techniques. The combination of high spa tial resolution using the methods presented in this article and the high temporal resolution of EPI opens opportunities for research into fMRl contrast mechanisms. Increase of percent signal change as the resolution increases is attributed to reduction of partial volume effects of activated voxels. Histo grams of fMRl pixel responses are progressively weighted to higher percent signal changes as the resolution increases. The conclusion has been reached that halfk‐space GR‐EPI is generally superior to fullk‐space GR‐EPI and should be used even for low‐resolution (64 × 64) EPI.

This publication has 21 references indexed in Scilit: