A mimic osseointegrated implant model for three‐dimensional finite element analysis
- 17 December 2002
- journal article
- research article
- Published by Wiley in Journal of Oral Rehabilitation
- Vol. 30 (1) , 41-45
Abstract
summary The purpose of this study was to develop a new three‐dimensional (3D) mimic model of an osseointegrated implant for finite element analysis (FEA) and to evaluate stress distributions in comparison with a model commonly used in most studies as a control. Based on the 3D computer graphic data obtained by serial in vivo bucco‐lingual peri‐implant bone structure at 75 μm interval in monkey, a mimic FEA model with trabecular structure and a control model with uniform cancellous bone were constructed. A vertical load of 143 N was applied at the top of the implant and induced stress was evaluated at the peri‐implant bone. In the mimic model, stress was distributed at both cortical and cancellous bones (1–5 MPa) in bucco‐lingual central planes, but concentrated at the cortical crest (3–7 MPa) in the mesio‐distal central plane. In contrast, the control model presented stress concentration at the cortical crest around the implant (5–14 MPa), with less stress (0–1 MPa) at the peri‐implant cancellous bone in both planes. The findings, that stress distribution at the peri‐implant bone were quite different between the mimic and control models, suggest the need to carefully interpret stress distribution in previous studies with models of uniform cancellous peri‐implant bone.Keywords
This publication has 11 references indexed in Scilit:
- The effectiveness of a new algorithm on a three-dimensional finite element model construction of bone trabeculae in implant biomechanicsJournal of Oral Rehabilitation, 1999
- The effectiveness of element downsizing on a three‐dimensional finite element model of bone trabeculae in implant biomechanicsJournal of Oral Rehabilitation, 1999
- A comparative evaluation of mandibular finite element models with different lengths and elements for implant biomechanicsJournal of Oral Rehabilitation, 1998
- 3D-FEA of osseointegration percentages and patterns on implant-bone interfacial stressesJournal of Dentistry, 1997
- Modeling the Mechanical Behavior of the Jaws and Their Related Structures By Finite Element (Fe) AnalysisCritical Reviews in Oral Biology & Medicine, 1997
- A Three‐Dimensional Finite Element Stress Analysis of Angled Abutments for an Implant Placed in the Anterior MaxillaJournal of Prosthodontics, 1995
- Location of implants in the interforaminal region of the mandible and the consequences for the design of the superstructureJournal of Oral Rehabilitation, 1994
- A three-dimensional, finite-element analysis of bone around dental implants in an edentulous human mandibleArchives of Oral Biology, 1993
- The three-dimensional bone interface of an osseointegrated implant: A method for studyThe Journal of Prosthetic Dentistry, 1992
- Direct bone anchorage of dental implantsThe Journal of Prosthetic Dentistry, 1983