X-Ray Spectrometer with a Submicron X-Ray Beam for Ulsi Microanalysis
Authors:
- 1 January 1994
- journal article
- Published by Springer Nature in MRS Proceedings
- Vol. 338
Abstract
A fluorescent and diffraction X-ray spectrometer with a 0.8 μmϕ Xray beam has been developed. It allows simultaneous measurement of local strains and minute amounts of metal contaminants in fine ULSI devices. The minimum sample size for X-ray diffraction measurement with it is a 0.3 μm diameter by 0.2 μm deep volume. It is applied to analyze strain in Al lines, showing that strain in Al single-layer lines (no barrier) increases significantly as the line width is reduced below 1.5 μm. Introducing barrier metals reduces this dependence of strain on line width. It is also found that hillock and void formation has a very strong correlation to strain.Keywords
This publication has 6 references indexed in Scilit:
- Reaction Growth and Morphology of An Aluminide Compound in Al-Cu/Ti-W BilayersMRS Proceedings, 1992
- Reliability, Properties and Microstructure of W80Ti20 Diffusion Barriers Between Al-Based Metallizations and Si.MRS Proceedings, 1992
- Triangular Shaped Beam Technique in EB Exposure System EX-7 for ULSI Pattern FormationJapanese Journal of Applied Physics, 1989
- Residual Stress1987
- Determination of Strain and Stress Fields by Diffraction Methods1987
- X-Ray MicroscopyPhysics Today, 1961