On the Significance of Systolic Flow Waveform on Aortic Valve Energy Loss
- 20 July 2018
- journal article
- research article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 46 (12) , 2102-2111
Abstract
This study aims to quantitatively and qualitatively assess energy dissipation in the aortic valve as a function of systolic aortic flow waveform representing pathologies where flow time-to-peak is delayed. A bioprosthetic valve was tested in the aortic position of a left-heart simulator under physiological pressure and flow conditions. The flow loop piston pump was programmed to generate three different flow waveforms each with a different peak time annotated as early peak (EP) with a rapid acceleration, mid peak (MP) and late peak (LP) with a rapid deceleration. Energy dissipation was calculated from flow and pressure measurements while sinus vorticity dynamics were evaluated using time-resolved planar particle image velocimetry. Average pressure gradients during systole are found 30.2 ± 0.19, 30.7 ± 0.25 and 32.9 ± 0.29 mmHg and average dissipation over systole is found 0.95 ± 0.026, 1.05 ± 0.034 and 1.25 ± 0.043 W for EP, MP and LP respectively. As systole’s acceleration phase is slower, sinus vortices are more likely to form, necessitating more energy exchange from shear layers inducing more viscous dissipation. EP found in healthy individuals is superior in terms of reducing energy dissipation and increasing aortic valve efficiency. In the context of possible left ventricular dysfunction and aortic stenosis, this means that delayed time-to-peak in the aortic flow waveform seen is not compensatory.Keywords
Funding Information
- National Institutes of Health (R01HL119824)
This publication has 25 references indexed in Scilit:
- Impact of patient-specific morphologies on sinus flow stasis in transcatheter aortic valve replacement: An in vitro studyThe Journal of Thoracic and Cardiovascular Surgery, 2019
- Stented valve dynamic behavior induced by polyester fiber leaflet material in transcatheter aortic valve devicesJournal of the Mechanical Behavior of Biomedical Materials, 2018
- Implantation Depth and Rotational Orientation Effect on Valve-in-Valve Hemodynamics and Sinus FlowThe Annals of Thoracic Surgery, 2018
- Effect of severe bioprosthetic valve tissue ingrowth and inflow calcification on valve-in-valve performanceJournal of Biomechanics, 2018
- Aortic sinus flow stasis likely in valve-in-valve transcatheter aortic valve implantationThe Journal of Thoracic and Cardiovascular Surgery, 2017
- On the Mechanics of Transcatheter Aortic Valve ReplacementAnnals of Biomedical Engineering, 2016
- Development of the trailing shear layer in a starting jet during pinch-offJournal of Fluid Mechanics, 2012
- The Cardiac Cycle and the Physiologic Basis of Left Ventricular Contraction, Ejection, Relaxation, and FillingHeart Failure Clinics, 2008
- Vorticity dynamics of a bileaflet mechanical heart valve in an axisymmetric aortaPhysics of Fluids, 2007
- A universal time scale for vortex ring formationJournal of Fluid Mechanics, 1998