Correlation of the clinical and physical image quality in chest radiography for average adults with a computed radiography imaging system
Authors:
- C S MooreCMC S Moore
- T J WoodTWT J Wood
- A W BeavisABA W Beavis
- J R SaundersonJSJ R Saunderson
- 1 July 2013
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 86 (1027) , 20130077
Abstract
The purpose of this study was to examine the correlation between the quality of visually graded patient (clinical) chest images and a quantitative assessment of chest phantom (physical) images acquired with a computed radiography (CR) imaging system. The results of a previously published study, in which four experienced image evaluators graded computer-simulated postero-anterior chest images using a visual grading analysis scoring (VGAS) scheme, were used for the clinical image quality measurement. Contrast-to-noise ratio (CNR) and effective dose efficiency (eDE) were used as physical image quality metrics measured in a uniform chest phantom. Although optimal values of these physical metrics for chest radiography were not derived in this work, their correlation with VGAS in images acquired without an antiscatter grid across the diagnostic range of X-ray tube voltages was determined using Pearson's correlation coefficient. Clinical and physical image quality metrics increased with decreasing tube voltage. Statistically significant correlations between VGAS and CNR (R=0.87, p<0.033) and eDE (R=0.77, p<0.008) were observed. Medical physics experts may use the physical image quality metrics described here in quality assurance programmes and optimisation studies with a degree of confidence that they reflect the clinical image quality in chest CR images acquired without an antiscatter grid. A statistically significant correlation has been found between the clinical and physical image quality in CR chest imaging. The results support the value of using CNR and eDE in the evaluation of quality in clinical thorax radiography.Keywords
This publication has 38 references indexed in Scilit:
- Suspension criteria for image monitors and viewing boxesRadiation Protection Dosimetry, 2012
- A method to produce and validate a digitally reconstructed radiograph-based computer simulation for optimisation of chest radiographs acquired with a computed radiography imaging systemBritish Journal of Radiology, 2011
- Effective dose efficiency: an application-specific metric of quality and dose for digital radiographyPhysics in Medicine & Biology, 2011
- Detector or System? Extending the Concept of Detective Quantum Efficiency to Characterize the Performance of Digital Radiographic Imaging SystemsRadiology, 2008
- Investigation of optimum X-ray beam tube voltage and filtration for chest radiography with a computed radiography systemBritish Journal of Radiology, 2008
- Optimisation of image plate radiography with respect to tube voltageRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: effect of anatomical noiseRadiation Protection Dosimetry, 2005
- Nodule detection in digital chest radiography: part of image background acting as pure noiseRadiation Protection Dosimetry, 2005
- A method for measuring the presampled MTF of digital radiographic systems using an edge test deviceMedical Physics, 1998
- Beam quality independent attenuation phantom for estimating patient exposure from x-ray automatic exposure controlled chest examinationsMedical Physics, 1984