Wide-range load sensor using quartz crystal resonator for biological signal detection
Authors:
- 1 May 2014
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 4405-4410
Abstract
A load sensor with high sensitivity, a wide measurement range, and a small size was developed by using an AT-cut quartz crystal resonator (QCR). The quartz crystal generates a charge that is proportional to the external force. Because it has high sensitivity and excellent temperature stability, it has been used for various sensors. In particular, a QCR has an inherently superior static-load-sensing characteristic. However, a QCR is fragile and easily broken by a stress concentration. Moreover, a retention mechanism is required to efficiently transmit the load, and it is necessary to fix the QCR firmly to avoid a horizontal force. Moreover, it is very difficult to miniaturize the retention mechanism because the fabrication and assembly process is complicated. We previously proposed a miniaturized sensor element that was developed using microfabrication. The QCR load sensor had an enormously wide force-sensing range of greater than 104 N. However, the output was easily affected by a change in the parasitic capacitance around the QCR. The objective of this study was to improve the load-measurement resolution and stabilize the sensor output for application to biological signal detection. We fabricated a QCR sensor with a sensitivity of 973 Hz/N and succeeded in detecting multiple biological signals (respiration, heartbeat, and posture) with using proposed QCR load sensor.Keywords
This publication has 12 references indexed in Scilit:
- Design, fabrication and characterization of compact force sensor using AT-cut quartz crystal resonators2008
- Piezoelectric Vibration-Type Tactile Sensor Using Elasticity and Viscosity Change of StructureIEEE Sensors Journal, 2007
- DNA sensing with a quartz crystal device for determination of microorganismsSensors and Actuators B: Chemical, 2007
- A thickness-shear quartz resonator force sensor with dual-mode temperature compensationIEEE Sensors Journal, 2003
- Some investigations on doubly-rotated quartz resonant temperature sensorsSensors and Actuators A: Physical, 2002
- Review of thickness-shear mode quartz resonator sensors for temperature and pressureIEEE Sensors Journal, 2001
- Detection of volatile amines using a quartz crystal with gold electrodesSensors and Actuators B: Chemical, 1999
- Characteristics of a Rectangular AT Cut Quartz Resonator as a Force SensorTransactions of the Society of Instrument and Control Engineers, 1996
- Frequency-output force sensor using a multimode doubly rotated quartz resonatorSensors and Actuators A: Physical, 1995
- Force-Frequency Coefficient of Singly Rotated Vibrating Quartz CrystalsIBM Journal of Research and Development, 1968