The effect of syringe size on the performance of an infusion pump
Authors:
- David J. StewardDSDavid J. Steward
- 1 July 1999
- journal article
- research article
- Published by Wiley in Pediatric Anesthesia
- Vol. 9 (4) , 335-337
Abstract
Summary: This study investigated the effect of using three different sized syringes on the accuracy of fluid delivery when using an infusion pump at low infusion rates (1 ml·h−1). The study also measured the influence of the syringe size on the time to occlusion alarm, and on the size of the subsequent bolus dose that might be infused after relief of the occlusion which triggered the alarm. The use of a larger size syringe was found not to affect the accuracy of infusate delivery, but delayed the time to occlusion alarm and increased the size of the postocclusion bolus dose.Keywords
This publication has 3 references indexed in Scilit:
- Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectorsIntensive Care Medicine, 1997
- Syringe Pumps for Infusion of Vasoactive DrugsAnesthesia & Analgesia, 1994
- The accuracy and variability of bolus injections with different sized syringesCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1992