Safety of endoscopic procedures with monopolar versus bipolar instruments in an ex vivo porcine model
Authors:
Open Access
- 31 January 2020
- journal article
- research article
- Published by Springer Nature in BMC Gastroenterology
- Vol. 20 (1) , 1-6
Abstract
Background: Monopolar instruments are generally used in colorectal endoscopic mucosal resection (EMR). Bipolar instruments have previously been reported to be as safe as monopolar instruments. We sought to compare the safety of the monopolar and bipolar snare and hemostatic forceps in an animal model. Methods: We created 5-mm, 10-mm, and 15-mm target lesions on an ex vivo porcine rectum. Two lesions of each size were resected via monopolar polypectomy (M-P), monopolar EMR (M-E), bipolar polypectomy (B-P), and bipolar EMR (B-E). We performed a pathological evaluation of the conditions of perforation and the effects of burning on the tissues. In addition, we burned the muscularis propria covered with submucosal layer using monopolar and bipolar hemostatic forceps and performed pathological evaluations. Results: Polypectomy and EMR were performed in a total of 24 target lesions. A perforation was found on histology in one case of M-P and one case of M-E after removing target lesions of 15 mm in diameter. There were no perforations during endoscopic resection using the bipolar snare. The thermal denaturation in B-P did not reach the muscularis propria layer regardless of the size of the target lesion. Although thermal damage after using monopolar hemostatic forceps was extensive, thermal denaturation was only seen on the surface of the submucosal layer when bipolar hemostatic forceps were used. Conclusions: Bipolar instruments cause less damage to the tissue than monopolar instruments. Our results also suggest that bipolar instruments may be safer than monopolar instruments in endoscopic procedures for colorectal lesions.Keywords
This publication has 30 references indexed in Scilit:
- Colonoscopic Polypectomy and Long-Term Prevention of Colorectal-Cancer DeathsThe New England journal of medicine, 2012
- Safety of cold polypectomy for < 10 mm polyps at colonoscopy: a prospective multicenter studyEndoscopy, 2011
- Evaluation of complications related to therapeutic colonoscopy using the bipolar snareSurgical Endoscopy, 2011
- Endoscopic hemostatic devicesGastrointestinal Endoscopy, 2009
- Routine coagulation of visible vessels may prevent delayed bleeding after endoscopic submucosal dissection - An analysis of risk factorsEndoscopy, 2008
- Risk factors for severe delayed postpolypectomy bleedingEndoscopy, 2008
- Polypectomy devicesGastrointestinal Endoscopy, 2007
- Postpolypectomy lower GI bleeding: Descriptive analysisGastrointestinal Endoscopy, 2000
- Prevention of Colorectal Cancer by Colonoscopic PolypectomyThe New England journal of medicine, 1993