Electrode array design determines scalar position, dislocation rate and angle and postoperative speech perception
Open Access
- 15 November 2021
- journal article
- research article
- Published by Springer Nature in European Archives of Oto-Rhino-Laryngology
- Vol. 279 (9) , 4257-4267
Abstract
Purpose: The aim of this study is to examine the scalar dislocation rate in straight and perimodiolar electrode arrays in relation to cochlear morphology. Furthermore, we aim to analyze the specific dislocation point of electrode arrays depending on their design and shape and to correlate these results to postoperative speech perception. Methods: We conducted a comparative analysis of patients (ears: n = 495) implanted between 2013 and 2018 with inserted perimodiolar or straight electrode arrays from Cochlear™ or MED-EL. CBCT (cone beam computed tomography) was used to determine electrode array position (scalar insertion, intra-cochlear dislocation, point of dislocation and angular insertion depth). Furthermore, cochlear morphology was measured. The postoperative speech discrimination was compared regarding electrode array dislocation, primary scalar insertion and angular insertion depth. Results: The electrode array with the highest rate of primary SV insertions was the CA; the electrode array with the highest rate of dislocations out of ST was the FlexSoft. We did not find significantly higher dislocation rates in cochleostomy-inserted arrays. The angle of dislocation was electrode array design-specific. A multivariate nonparametric analysis revealed that the dislocation of the electrode array has no significant influence on postoperative speech perception. Nevertheless, increasing angular insertion depth significantly reduced postoperative speech perception for monosyllables. Conclusion: This study demonstrates the significant influence of electrode array design on scalar location, dislocation and the angle of dislocation itself. Straight and perimodiolar electrode arrays differ from each other regarding both the rate and place of dislocation. Insertion via cochleostomy does not lead to increased dislocation rates in any of the included electrode arrays. Furthermore, speech perception is significantly negatively influenced by angular insertion depth.Keywords
Funding Information
- Universitätsklinikum Freiburg
This publication has 31 references indexed in Scilit:
- Factors Affecting Open-Set Word Recognition in Adults With Cochlear ImplantsEar & Hearing, 2013
- Systemic Immunity Influences Hearing Preservation in Cochlear ImplantationOtology & Neurotology, 2012
- Insertionsergebnisse von Contour™- und Contour-Advance™-ElektrodenHNO, 2011
- The Internal Dimensions of the Cochlear Scalae With Special Reference to Cochlear Electrode Insertion TraumaOtology & Neurotology, 2010
- Considerations for design of future cochlear implant electrode arrays: Electrode array stiffness, size,Journal of Rehabilitation Research and Development, 2008
- Role of Electrode Placement as a Contributor to Variability in Cochlear Implant OutcomesOtology & Neurotology, 2008
- Quality Control After Insertion of the Nucleus Contour and Contour Advance Electrode in AdultsEar & Hearing, 2007
- The Size of the Cochlea and Predictions of Insertion Depth Angles for Cochlear Implant ElectrodesAudiology and Neurotology, 2006
- Evaluation of the Advance Off-Stylet Insertion Technique and the Cochlear Insertion Tool in Temporal BonesOtology & Neurotology, 2005
- Comparative Study of Cochlear Damage With Three Perimodiolar Electrode DesignsThe Laryngoscope, 2003