Flush-mounted probes in the divertor plates of Tokamak de Varennes
Authors:
- 1 January 1995
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (1) , 154-159
Abstract
Arrays of 12 flush-mounted Langmuir probes have been installed in both the upper and lower outboard divertor plates of TdeV (Tokamak de Varennes). Measurements of electron temperature and density are obtained using a novel method of analysis which assumes that charge collection is dominated by the finite size of the electrostatic sheath. The technique is validated by comparing measurements from a cylindrical and a flush probe which are operated simultaneously in the same plasma.Keywords
This publication has 6 references indexed in Scilit:
- Sheath structure in a magnetized plasmaPhysics of Fluids B: Plasma Physics, 1993
- Retention of intrinsic gases in TdeV's biasable divertorJournal of Nuclear Materials, 1992
- Gundestrup: A Langmuir/Mach probe array for measuring flows in the scrape-off layer of TdeVReview of Scientific Instruments, 1992
- Investigation of the fluxes to a surface at grazing angles of incidence in the tokamak boundaryPlasma Physics and Controlled Fusion, 1990
- Electric probes in plasmasJournal of Vacuum Science & Technology A, 1986
- Plasma–wall transition in an oblique magnetic fieldThe Physics of Fluids, 1982