Dynamic Test to Diagnose Loose Uncemented Femoral Total Hip Components
Authors:
- Fred FletcherFFFred Fletcher
- Thomas DonaldsonTDThomas Donaldson
- Ray WasielewskiRWRay Wasielewski
- Mark PetersonMPMark Peterson
- Harry RubashHRHarry Rubash
- 1 September 1996
- journal article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 330 (330) , 115-123
Abstract
A new computed tomographic rotation study was developed to determine the stability of uncemented femoral components after total hip arthroplasty. The computed tomographic rotation study measures femoral component version with the leg in maximum external and internal rotation. The component is considered rotationally unstable, and thus loose, if these angles differ by more than 2 degrees. The results of the computed tomographic rotation study were confirmed with intraoperative inspection in 50 patients: 13 patients (Group A) with definite radiographic loosening, 22 patients (Group B) with equivocal radiographic loosening, and 15 patients (Group C) with documented stability. The results of the computed tomographic rotation study agreed with intraoperative findings for all patients in Groups A and C, and 20 of 22 patients in Group B (1 false negative and 1 false positive). Statistical analysis of Group B showed that the computed tomographic rotation study had a sensitivity of 91.6% and a specificity of 90.0%. The positive predictive value was 91.6%. These results indicate that the computed tomographic rotation study can supplement equivocal radiographic findings in assessing loosening of an uncemented femoral component.Keywords
This publication has 14 references indexed in Scilit:
- Technique to detect migration of femoral components of total hip arthroplasties on conventional radiographsThe Journal of Arthroplasty, 1991
- Arthroscintography for detection of femoral component looseningThe Journal of Arthroplasty, 1991
- Endosteal erosion in association with stable uncemented femoral components.Journal of Bone and Joint Surgery, 1990
- Porous-coated anatomic total hip prostheses: radiographic analysis and clinical correlation.Radiology, 1990
- Telemetric force measurements across the hip after total arthroplasty.Journal of Bone and Joint Surgery, 1988
- Femoral Neck Torsion Angle Measurement by Computed TomographyJournal of Computer Assisted Tomography, 1987
- Definition of endoprosthetic loosening: Comparison of arthrography, scintigraphy and roentgen stereophotogrammetry in prosthetic hipsActa Orthopaedica, 1985
- Failed total hip replacementJournal of Bone and Joint Surgery, 1984
- Prosthetic hip replacements: plain films and arthrography for component looseningAmerican Journal of Roentgenology, 1982
- Fractured Stem in Total Hip Replacement:A Clinical Review of 120 CasesActa Orthopaedica, 1982