Rapid Communication : Effects of Steris 1 ™ Sterilization and Cidex ® Ortho-Phthalaldehyde High-Level Disinfection on Durability of New-Generation Flexible Ureteroscopes
- 1 September 2007
- journal article
- research article
- Published by SAGE Publications in Journal of Endourology
- Vol. 21 (9) , 985-992
Abstract
Background and Purpose: The effects of commonly used reprocessing methods on flexible ureteroscope longevity have never been examined. We prospectively studied the effects of Steris 1™ sterilization and Cidex® ortho-phthalaldehyde (OPA) high-level disinfection (HLD) on the image quality, physical structure, and deflective properties of two new flexible ureteroscopes. Materials and Methods: Two identical “out-of-the-box” Storz 11278AU1 flexible ureteroscopes (Karl Storz Endoscopy, Tuttlingen, Germany) were sterilized individually using the Steris 1™ system (Steris Mentor, Ohio) or disinfected with Cidex® OPA (Advanced Sterilization Products, J&J, Irvine, CA) for 100 trials followed by a crossover to the other method for another 100 trials over a period of 1 year. After every five trials, optical quality, angle of deflection, and fiber damage were analyzed in the laboratory. Throughout the study, neither of these ureteroscopes was used clinically. Results: After 100 trials, ureteroscope 1, which was sterilized initially in the Steris system, had a 12-mm tear on its shaft (noted after the 17th trial), 297 damaged fibers, and a 37% drop in resolution (loss of 3.75 lines/mm). There was no change in deflection from baseline. In contrast, after 100 cycles, ureteroscope 2, which was subjected to HLD with Cidex® OPA, had no visible external damage, a 0% change in resolution, 10 damaged fibers, and no change in deflection. After the crossover, ureteroscope 2 developed a semilunar defect that obscured the endoscopic view, whereas there was no further significant damage to ureteroscope 1. Conclusion: After 100 cycles, the Steris 1 system rendered the flexible ureteroscope unusable, whereas HLD with Cidex® OPA had minimal adverse impact.Keywords
This publication has 26 references indexed in Scilit:
- Frequency of Ureteroscope Damage Seen at a Tertiary Care CenterJournal of Urology, 2006
- Durability of Flexible Ureteroscopes: A Randomized, Prospective StudyJournal of Urology, 2006
- COMPARISON OF FLEXIBLE URETEROSCOPES: DEFLECTION, IRRIGANT FLOW AND OPTICAL CHARACTERISTICSJournal of Urology, 2005
- Performance and Durability of Leading Flexible UreteroscopesJournal of Endourology, 2004
- Reprocessing endoscopes: United States perspectiveJournal of Hospital Infection, 2004
- Two cases of ortho-phthalaldehyde induced allergic reactions in patients undergoing surveillance cystoscopyJournal of Allergy and Clinical Immunology, 2004
- Does the Cleaning Technique Influence the Durability of the <9F Flexible Ureteroscope?Journal of Endourology, 2001
- Efficacy of high-level disinfectants for reprocessing GI endoscopes in simulated-use testingGastrointestinal Endoscopy, 2001
- In-hospital evaluation of orthophthalaldehyde as a high level disinfectant for flexible endoscopesJournal of Hospital Infection, 1994
- Clinical Application of Fiber-Optic PyeloureteroscopeUrologia Internationalis, 1971