Abstract
This study develops a three-dimensional finite element torso model with bidomain myocardium to simulate the transmembrane potential (TMP) of the heart induced by defibrillation fields. The inhomogeneities of the torso are modeled as eccentric spherical volumes with both the curvature and the rotation features of cardiac fibers incorporated in the myocardial region. The numerical computation of the finite element bidomain myocardial model is validated by a semianalytic solution. The simulations show that rotation of fiber orientation through the depth of the myocardial wall changes the pattern of polarization and decreases the amount of cardiac tissue polarized compared to the idealized analytic model with no fiber rotation incorporated. The TMP induced by transthoracic and transvenous defibrillation fields are calculated and visualized. The TMP is quantified by a continuous measure of the percentage of myocardial mass above a potential gradient threshold. Using this measure, the root mean square differences in TMP distribution produced by reversing the electrode polarity for anterior-posterior and transvenous electrode configurations are 13.6 and 28.6%, respectively. These results support the claim that a bidomain model of the heart predicts a change of defibrillation threshold with reversed electrode polarity.Keywords
This publication has 27 references indexed in Scilit:
- Polarity Reversal Improves Defibrillation Efficacy in Patients Undergoing Transvenous Cardioverter Defibrillator Implantation with Biphasic SbocksPacing and Clinical Electrophysiology, 1997
- The Effect of Polarity of the Initial Phase of a Biphasic Shock Waveform on the Defibrillation Threshold of Pectorally Implanted DefibrillatorsPacing and Clinical Electrophysiology, 1997
- Effects of Waveform and Polarity on Defibrillation Thresholds in Humans Using a Transvenous Lead SystemThe American Journal of Cardiology, 1996
- Effects of Initial Polarity on Defibrillation Threshold with Biphasic PulsesPacing and Clinical Electrophysiology, 1995
- Effect of shock polarity on ventricular defibrillation threshold using a transvenous lead systemJACC, 1994
- Successful implantation of cardiovascular-defibrillator systems in patients with elevated defibrillation thresholdsJACC, 1993
- The automatic implantable cardioverter-defibrillator: Effect of patch polarity on defibrillation thresholdJACC, 1991
- Potential distribution in three-dimensional periodic myocardium. I. Solution with two-scale asymptotic analysisIEEE Transactions on Biomedical Engineering, 1990
- Evaluation of electrode polarity on defibrillation efficacyThe American Journal of Cardiology, 1989
- The specific resistance of biological material—A compendium of data for the biomedical engineer and physiologistMedical & Biological Engineering & Computing, 1967