Temporary Right Ventricular Mechanical Support in High‐Risk Left Ventricular Assist Device Recipients Versus Permanent Biventricular or Total Artificial Heart Support
- 4 April 2013
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 37 (6) , 523-530
Abstract
Early planned institution of temporary right ventricular assist device (RVAD) support with the CentriMag (Levitronix LLC, Waltham, MA, USA) in left ventricular assist device (LVAD) recipients was compared with permanent biventricular assist device (BVAD) or total artificial heart (TAH) support. Between 2007 and 2011, 77 patients (age range: 25–70 years) with preoperative evidence of biventricular dysfunction (University of Pennsylvania score >50; University of Michigan score >5) were included. Forty‐six patients (38 men; median age 54.5 years, range: 25–70 years) underwent LVAD placement combined with temporary RVAD support (group A); in 31 patients (25 men; median age 56.7 years, range: 28–68 years), a permanent BVAD or TAH implantation (group B) was performed. Within 30 days, 12 patients from group A (26.08%) and 14 patients from group B (45.1%) died on mechanical support (P = 0.02). Thirty patients (65.2%) in group A were weaned from temporary RVAD support and three (6.5%) underwent permanent RVAD (HeartWare, Inc., Framingham, MA, USA) placement. A total of 26 patients (56.5%) were discharged home in group A versus 17 (54.8%) in group B (P = 0.56). Three patients (8.5%) received heart transplantation in group A and six (19.3%) in group B (P = 0.04). In group A, 90‐day and 6‐month survival was 54.3% (n = 25) versus 51.6% (n = 16) in group B (P = 0.66). In group A, 1‐year survival was 45.6% (n = 21) versus 45.1% (n = 14) in group B (P = 0.81). The strategy of planned temporary RVAD support in LVAD recipients showed encouraging results if compared with those of a similar permanent BVAD/TAH population. Weaning from and removal of the temporary RVAD support may allow patients to be on LVAD support only despite preoperative biventricular dysfunction.Keywords
This publication has 26 references indexed in Scilit:
- Levitronix CentriMag Third-Generation Magnetically Levitated Continuous Flow Pump as Bridge to SolutionASAIO Journal, 2011
- Tricuspid valve repair with left ventricular assist device implantation: Is it warranted?The Journal of Heart and Lung Transplantation, 2011
- Advances in mechanical circulatory support: Year in reviewThe Journal of Heart and Lung Transplantation, 2011
- Total ventricular assist for long-term treatment of heart failureThe Journal of Thoracic and Cardiovascular Surgery, 2011
- Right-to-left ventricular end-diastolic diameter ratio and prediction of right ventricular failure with continuous-flow left ventricular assist devicesThe Journal of Heart and Lung Transplantation, 2011
- Risk Factors Predictive of Right Ventricular Failure After Left Ventricular Assist Device ImplantationThe American Journal of Cardiology, 2010
- Early planned institution of biventricular mechanical circulatory support results in improved outcomes compared with delayed conversion of a left ventricular assist device to a biventricular assist deviceThe Journal of Thoracic and Cardiovascular Surgery, 2009
- Risk Score Derived from Pre-operative Data Analysis Predicts the Need for Biventricular Mechanical Circulatory SupportThe Journal of Heart and Lung Transplantation, 2008
- The Right Ventricular Failure Risk ScoreJACC, 2008
- Cardiac Replacement with a Total Artificial Heart as a Bridge to TransplantationNew England Journal of Medicine, 2004