Miniature Fabry–Perot Fiber Optic Sensor for Intravascular Blood Temperature Measurements
Authors:
- 6 March 2013
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 13 (6) , 2155-2160
Abstract
This paper presents a miniature fiber optic sensor for intravascular blood temperature measurements. Based on the Fabry–Perot interferometric principle, the sensor is fabricated by using chemical etching and thermal deposition. Before the animal test, static temperature calibration experiments are performed to characterize the sensor's sensitivity, linearity, and hysteresis. A swine model is selected to perform the in-vivo experiment. During the in-vivo experiment, the swine intravascular blood temperature is measured at different locations in coronary arteries to demonstrate the sensor's performance. The sensor's usage in the thermal angioplasty application is successfully presented by capturing the rise and drop of local intravascular blood temperature variations.Keywords
This publication has 12 references indexed in Scilit:
- An experimental study on the concrete hydration process using Fabry–Perot fiber optic temperature sensorsMeasurement, 2012
- An ultra-fast fiber optic pressure sensor for blast event measurementsMeasurement Science and Technology, 2012
- A Miniature Fiber Optic Refractive Index Sensor Built in a MEMS-Based MicrochannelSensors, 2011
- In vitro and in vivo studies on thermistor‐based intracoronary temperature measurements: Effect of pressure and flowCatheterization and Cardiovascular Interventions, 2008
- Heart Disease and Stroke Statistics—2007 UpdateCirculation, 2007
- Increased local temperature in human coronary atherosclerotic plaques: an independent predictor of clinical outcome in patients undergoing a percutaneous coronary interventionJACC, 2001
- Reproducible optical fiber tips for photon scanning tunneling microscopy with very small (>5°) cone angleJournal of Lightwave Technology, 1998
- Thermal laser probe angioplasty: Influence of constant tip temperature, plaque composition, and probe/vessei diameter ratioLasers in Surgery and Medicine, 1990
- Transluminal Treatment of Arteriosclerotic ObstructionRadiology, 1989
- Balloon catheter technique for dilatation of constricted cerebral arteries after aneurysmal SAHActa Neurochirurgica, 1984