Physical characteristics of low and medium energy collimators for 123I imaging and simultaneous dual-isotope imaging
Authors:
- Y InoueY Inoue
- I ShirouzuISI Shirouzu
- T MachidaTMT Machida
- Y YoshizawaYYY Yoshizawa
- F AkitaFAF Akita
- I DoiIDI Doi
- T WatadaniTWT Watadani
- M NodaMNM Noda
- K YoshikawaKYK Yoshikawa
- K OhtomoKOK Ohtomo
- 1 November 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Nuclear Medicine Communications
- Vol. 24 (11) , 1195-1202
Abstract
Septal penetration of high energy photons may degrade 123I images obtained with a low energy collimator. We evaluated the physical characteristics of a low energy, high resolution (LEHR) collimator, special LEHR (SLEHR) collimator, and medium energy collimator for 123I studies. The cross-talk of 99mTc and 123I into the 201Tl window was also examined. Sensitivity and spatial resolution were measured with each collimator. Point sources of 99mTc and 123I were imaged at various source-collimator distances using multiple energy windows, and the effects of collimator choice on energy spectrum and spatial distribution of photopeak counts were assessed. For 99mTc, both sensitivity and resolution were similar with the low energy collimators, and higher sensitivity and lower resolution were observed with the medium energy collimator. For 123I, the full width at tenth maximum was larger for 123I than for 99mTc when using the LEHR collimator. Acquisitions with multiple energy windows revealed severe degradation due to septal penetration in imaging 123I with the LEHR collimator, especially at short distances. The degradation was reduced with the SLEHR collimator and further with the medium energy collimator. In both 99mTc and 123I imaging, cross-talk into the 201Tl window was larger at shorter distances and the largest with the LEHR collimator. In conclusion, variation in collimator geometry causes differences in the effect of septal penetration on 123I images and in cross-talk into the 201Tl window. The SLEHR collimator may be suitable for use in high resolution 123I imaging and simultaneous 99mTc/201Tl imaging. Use of the medium energy collimator appears to be preferable in quantitative 123I studies.Keywords
This publication has 8 references indexed in Scilit:
- Preliminary clinical results of photon energy recovery in simultaneous rest Tl-201/stress Tc-99m sestamibi myocardial SPECTJournal of Nuclear Cardiology, 2001
- Influence of high-energy photons on the spectrum of iodine-123 with low- and medium-energy collimators: consequences for imaging with 123 I-labelled compounds in clinical practiceEuropean Journal of Nuclear Medicine and Molecular Imaging, 1999
- Phantom evaluation of simultaneous thallium-201/technetium-99m aquisition in single-photon emission tomographyEuropean Journal of Nuclear Medicine and Molecular Imaging, 1996
- Optimal collimator choice for sequential iodine-123 and technetium-99m imagingEuropean Journal of Nuclear Medicine and Molecular Imaging, 1996
- Simultaneous Tc-99m/Tl-201 imaging using energy-based estimation of the spatial distributions of contaminant photonsIEEE Transactions on Nuclear Science, 1995
- A method of simultaneous dual-radionuclide cardiac imaging with technetium 99m and thallium 201. I: Analysis of interradionuclide crossover and validation in phantomsJournal of Nuclear Cardiology, 1994
- Error of uptake in dual energy acquisition with201TI and123I labeled radiopharmaceuticalsEuropean Journal of Nuclear Medicine and Molecular Imaging, 1990
- Imaging 123I with a scintillation camera. A study of detection performance and quality factor conceptsPhysics in Medicine & Biology, 1977