Reliability of Zephyr Bioharness and Fitbit Charge Measures of Heart Rate and Activity at Rest, During the Modified Canadian Aerobic Fitness Test, and Recovery
- 1 February 2019
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Strength & Conditioning Research
- Vol. 33 (2) , 559-571
Abstract
Nazari, G, MacDermid, JC, Sinden, KE, Richardson, J, and Tang, A. Reliability of Zephyr Bioharness and Fitbit Charge measures of heart rate and activity at rest, during the modified Canadian Aerobic Fitness Test, and recovery. J Strength Cond Res 33(2): 559–571, 2019—The purpose of this study was to determine the intrasession and intersession reliability of Zephyr Bioharness (ZB) and Fitbit Charge variables in both healthy men and women at rest, during the Modified Canadian Aerobic Fitness Test (mCAFT), and throughout recovery. Stratified convenience and snowball sampling were used to recruit 60 participants (30 women, 48 ± 15 years) and (30 men, 48 ± 15 years) from McMaster University student, staff, and faculty population. At rest, intrasession average heart rate (b·min−1). Intraclass correlation coefficients (ICCs) and Standard Error of Measurement [SEM] for Zephyr ranged from (0.94–0.97) [1.17–1.70] to (0.92–0.97) [1.45–2.10] for Fitbit Charge. During the mCAFT, the Zephyr ICCs and (SEM) ranged from 0.31–0.99 (1.28–8.10) to 0.45–0.99 (1.45–8.71) for the Fitbit Charge. Throughout the recovery, the ICCs and (SEM) ranged from 0.44–0.98 (1.26–10.47) to 0.45–0.98 (1.15–11.90) for Zephyr and Fitbit devices, respectively. At rest, intersession ICCs (SEM) for Zephyr and Fitbit ranged from 0.90–0.94 (1.73–2.37) to 0.88–0.94 (1.83–2.67), respectively. At mCAFT, the Zephyr ICCs (SEM) ranged from 0.91–0.97 (3.12–4.64) to 0.85–0.98 (3.28–4.88) for the Fitbit. Throughout the recovery, the ICCs (SEM) ranged from 0.93–0.97 (2.65–4.66) to 0.76–0.91 (3.17–4.67) for Zephyr and Fitbit devices, respectively. To conclude, both the ZB and Fitbit Charge devices demonstrated excellent reliability measures throughout the 3 phases. The findings from our study add to the existing pool of literature regarding the reliability parameters of wearable devices and suggest that stable and consistent measures of heart rate and physical activity can be obtained using ZB and Fitbit Charge devices among healthy male and female participants at rest, during a standardized submaximal fitness test (mCAFT), and throughout recovery.Keywords
This publication has 16 references indexed in Scilit:
- Fitbit®: An accurate and reliable device for wireless physical activity trackingInternational Journal of Cardiology, 2015
- Concurrent Validity of the Armour39 Heart Rate Monitor StrapJournal of Strength & Conditioning Research, 2014
- Revitalizing a Vital Sign: Improving Detection of Tachypnea at Primary TriageAnnals of Emergency Medicine, 2013
- Health care costs of physical inactivity in Canadian adultsApplied Physiology, Nutrition, and Metabolism, 2012
- Evidence-based intervention in physical activity: lessons from around the worldThe Lancet, 2012
- Correlation, Agreement, and Bland–Altman Analysis: Statistical Analysis of Method Comparison StudiesAmerican Journal of Ophthalmology, 2009
- When to use agreement versus reliability measuresJournal of Clinical Epidemiology, 2006
- Heart Rate MonitoringSports Medicine, 2003
- Simultaneous monitoring of heart rate and motion to assess energy expenditureMedicine & Science in Sports & Exercise, 1997
- A Study to Validate the Modified Canadian Aerobic Fitness TestCanadian Journal of Applied Physiology, 1995