Motion Artifact Reduction and Data Compression of Photoplethysmo-graphic Signals utilizing Cycle by Cycle Fourier Series Analysis
- 1 May 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10915281,p. 176-179
Abstract
Artifact-free clean photoplethysmographic (PPG) signals obtained at red and infrared (IR) wavelengths are required for the estimation of the level of oxygen saturation (SpO2) in arterial blood of a patient. Movement of a patient corrupts a PPG signal with motion artifacts and introduces large errors in the computation of SpO2- A novel method to remove motion artifacts from corrupted PPG signals by applying Fourier series analysis on a cycle by cycle basis is presented in this paper. Over and above artifact reduction, the proposed method also provides data compression. Experimental results indicate that the proposed method is insensitive to heart rate variation, introduces negligible error in the processed PPG signals due to the additional processing, preserves all the morphological features of the PPG, provides 35 dB reduction in motion artifact and achieves a data compression factor of 12.Keywords
This publication has 9 references indexed in Scilit:
- Virtual Instrument for the Measurement of Haemo-dynamic Parameters Using Photoplethysmograph2007 IEEE Instrumentation & Measurement Technology Conference IMTC 2007, 2006
- Motion Artifact Reduction in Photoplethysmography Using Independent Component AnalysisIEEE Transactions on Biomedical Engineering, 2006
- Comparison of wavelet transformation and adaptive filtering in restoring artefact-induced time-related measurementBiomedical Signal Processing and Control, 2006
- Motion Artifact In Pulse Oximetry2005
- Reduction of motion artifacts from photoplethysmographic recordings using a wavelet denoising approach2004
- Design of filter to reject motion artifact of pulse oximetryComputer Standards & Interfaces, 2004
- Adaptive reduction of motion artifact from photoplethysmographic recordings using a variable step-size LMS filter2003
- Design of Pulse Oximeters1997
- THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPHAmerican Journal of Physiology-Legacy Content, 1938