Effect of glutaraldehyde based cross-linking on the viscoelasticity of mitral valve basal chordae tendineae
Open Access
- 13 July 2018
- journal article
- research article
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 17 (1) , 93
Abstract
Mitral valve failure can require repair or replacement. Replacement bioprosthetic valves are treated with glutaraldehyde prior to implantation. The aim of this study was to determine the changes in mechanical properties following glutaraldehyde fixation of mitral valve chordae. To investigate the impact of glutaraldehyde on mitral valve chordae, 24 basal chordae were dissected from four porcine hearts. Anterior and posterior basal (including strut) chordae were used. All 24 chordae were subjected to a sinusoidally varying load (mean level 2N, dynamic amplitude 2N) over a frequency range of 0.5–10 Hz before and after glutaraldehyde treatment. The storage and loss modulus of all chordal types decreased following glutaraldehyde fixation. The storage modulus ranged from: 108 to 119 MPa before fixation and 67.3–87.4 MPa following fixation for basal chordae; 52.3–58.4 MPa before fixation and 47.9–53.5 MPa following fixation for strut chordae. Similarly, the loss modulus ranged from: 5.47 to 6.25 MPa before fixation and 3.63–4.94 MPa following fixation for basal chordae; 2.60–2.97 MPa before fixation and 2.31–2.93 MPa following fixation for strut chordae. The viscoelastic properties of mitral valve chordae are affected by glutaraldehyde fixation; in particular, the reduction in storage moduli decreased with an increase in chordal diameter.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (M114612B)
- Arthritis Research UK (H0671)
This publication has 45 references indexed in Scilit:
- Mechanics of the Mitral Annulus in Chronic Ischemic CardiomyopathyAnnals of Biomedical Engineering, 2013
- Cardiac crossroads: deciding between mechanical or bioprosthetic heart valve replacementPatient Preference and Adherence, 2011
- Freeze-thaw treatment effects on the dynamic mechanical properties of articular cartilageBMC Musculoskeletal Disorders, 2010
- Viscoelastic properties of bovine articular cartilage attached to subchondral bone at high frequenciesBMC Musculoskeletal Disorders, 2009
- Heart valve function: a biomechanical perspectivePhilosophical Transactions of the Royal Society B: Biological Sciences, 2007
- Midterm Results After Stentless Mitral Valve ReplacementCirculation, 2003
- A Mechanism for the Decrease in Stiffness of Bioprosthetic Heart Valve Tissues After Cross-LinkingASAIO Journal, 1996
- Glutaraldehyde fixation alters the internal shear properties of porcine aortic heart valve tissueThe Annals of Thoracic Surgery, 1995
- Morphology of the Human Mitral ValveCirculation, 1970
- Morphology of the Human Mitral ValveCirculation, 1970