Abstract
Tracheal tube (TT) displacement during general anaesthesia may result in life-threatening complications and continuous direct vision of the position of the tube may enable safer management. The ETView tracheoscopic ventilation tube (TVT™) is a single-use TT incorporating a video camera and a light source in its tip. The view from the tip appears continuously on a portable monitor in the anaesthetist's vicinity. This study was designed to test the ETView TVT™ in monitoring the TT position during general anaesthesia. In this prospective study, the ETView TVT™ was used to ventilate the lungs of 30 adult patients undergoing percutaneous nephrolithotomy (PCNL), which required changing patient position three times. During surgery, the anaesthetist followed the carinal view on the ETView TVT™ portable monitor. Tube movement within 1 cm was recorded, as was the need for repositioning of the tube when the carina was not seen on the camera monitor. During anaesthesia, tiny movements synchronous with heart beats and lung ventilation were observed. Tube movement of 1 cm was detected in eight (26%) patients. In two (7%) patients, the carina was no longer viewed after moving to the lithotomy position and the tube was repositioned. None of the events was associated with changes in oxygen saturation, end-tidal CO2, or airway pressure. We found that the ETView TVT™ facilitated surveillance of tube position by providing a clear high-quality view of the carina, throughout PCNL with several changes of patient position.Keywords
This publication has 16 references indexed in Scilit:
- Mortality related to anaesthesia in France: analysis of deaths related to airway complications*Anaesthesia, 2009
- Trends in Anesthesia-related Death and Brain DamageAnesthesiology, 2006
- Identification of Endotracheal Tube Malpositions Using Computerized Analysis of Breath Sounds via Electronic StethoscopesAnesthesia & Analgesia, 2005
- A Preliminary Study of the Optimal Anesthesia Positioning for the Morbidly Obese PatientObesity Surgery, 2003
- Pneumoperitoneum as a Risk Factor for Endobronchial Intubation During Laparoscopic Gynecologic SurgeryAnesthesia & Analgesia, 1998
- Adverse Respiratory Events in Anesthesia: A Closed Claims AnalysisAnesthesiology, 1990
- Radiographic detection of esophageal malpositioning of endotracheal tubes.American Journal of Roentgenology, 1990
- Role of Monitoring Devices in Prevention of Anesthetic MishapsAnesthesiology, 1989
- Endobronchial IntubationAnesthesiology, 1987
- EFFECT OF THE TRENDELENBURG TILT AND OTHER PROCEDURES ON THE POSITION OF ENDOTRACHEAL TUBESThe Lancet, 1969