Abstract
Objective:Radiopaque markers are commonly deployed following breast biopsies to indicate the location of the targeted lesion. A frequently encountered complication is the displacement of these markers. This study compares the degree of displacement among four newer generation markers after stereotactic core needle biopsy.Materials:Eighty consecutive biopsies were performed at three breast center sites. The markers included: HydroMARK®, MammoMARK™, MammoStar™, and SecurMark®. Each marker was composed of a radiopaque core with a unique polymeric encasing component. Post-procedure mammograms were obtained and the degree of marker displacement was measured.Results:MammoMARK™ exhibited the greatest mean net displacement, followed by HydroMARK®, SecurMark®, and MammoStar™ (13.9, 7.7, 5.8, and 4.7 mm, respectively), though these differences did not reach statistical significance (p = 0.398). Seventy-three percent of the markers did not displace at all. However, in the 19 of 22 markers in which displacement occur...