A comparison of anaesthetic breathing systems during spontaneous ventilation

Authors:

Abstract
Summary: Five anaesthetic breathing systems (Magill, Lack, Humphrey ADE, enclosed Magill and Bain) were compared using spontaneous ventilation in a simple lung model. The fresh gas flow at which rebreathing occurred was determined for each system by the application of four modified definitions of rebreathing. Two were based on the measurement of minimum inspired and two on end‐expired carbon dioxide. The four A systems performed similarly with each individual definition. The rebreathing points found for each individual breathing system differed markedly between definitions, with those determined by the minimum inspired CO 2 , occurring at low, and probably misleading, FGF/V̇ E ratio. The Bain system demonstrated rebreathing at considerably higher fresh gas flows whichever definition was used.