Nanomechanical evaluation of nickel–titanium surface properties after alkali and electrochemical treatments
- 29 January 2008
- journal article
- Published by The Royal Society in Journal of The Royal Society Interface
- Vol. 5 (26) , 1009-1022
Abstract
In this paper, the suitability of alkali treatment followed by heat treatment at 600°C, and spark oxidation for nickel–titanium, intended for medical applications such as pins, wires and clamps, was evaluated on the basis of nanomechanical and wear testing. In addition, the chemical composition and topography of the surface layer, wetting ability, corrosion resistance and influence of the heat treatment on structure of the alloy were also investigated. The results showed that the highest hardness was observed for alkali-treated samples, and this could be correlated with the structure of the sample that contained martensite and a higher phase transformation temperature. This treatment caused a very large increase of nickel in the top layer and decreased resistance in pitting corrosion. These results disqualified the treatment to be considered as useful for medical applications. On the other hand, the hardness of the oxidized samples was at the same level as that obtained for ground reference samples. Moreover, the oxide layer was enriched with phosphorus, and it was predominantly composed of TiO 2 and phosphorus oxides. This 3.1 μm thick layer had good adhesion to the substrate as indicated by scratch testing and wear resistant in nanowear testing. However, the oxidation did not significantly increase the corrosion resistance of the alloy compared with reference samples.Keywords
This publication has 43 references indexed in Scilit:
- Hydroxyapatite/titania sol–gel coatings on titanium–zirconium alloy for biomedical applications☆Acta Biomaterialia, 2007
- Bioactive layer formation on alkaline-acid treated titanium in simulated body fluidMaterials & Design (1980-2015), 2007
- High Voltage Pulse Anodization of a NiTi Shape Memory AlloyJournal of the Electrochemical Society, 2007
- Deposition of TiN coatings on shape memory NiTi alloy by plasma immersion ion implantation and depositionThin Solid Films, 2006
- Oxidized, bioactive implants are rapidly and strongly integrated in bone. Part 1 – experimental implantsClinical Oral Implants Research, 2006
- Fabrication and characteristics of bioactive sodium titanate/titania graded film on NiTi shape memory alloyJournal of Biomedical Materials Research Part A, 2005
- Surface treatment of NiTi for medical applicationsMinimally Invasive Therapy & Allied Technologies, 2004
- Structure Modification of Surface Layers of Ti6Al4V ELI ImplantsKey Engineering Materials, 2003
- Apatite formation on surfaces of ceramics, metals and polymers in body environmentActa Materialia, 1998
- XPS characterisation of surface modified Ni-Ti shape memory alloyMaterials Science and Engineering: A, 1997