Comparison of Procedures for Evaluating Laboratory Performance in External Quality Assessment Schemes for Lead in Blood and Aluminum in Serum Demonstrates the Need for Common Quality Specifications
Open Access
- 1 November 2002
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 48 (11) , 2000-2007
Abstract
Background: The different scoring methods used by eight European External Quality Assessment Schemes (EQASs) for occupational and environmental laboratory medicine were compared to develop suitable quality specifications as a step toward harmonization. Methods: Real results for blood lead and serum aluminum assays, reported by participants in Italian and United Kingdom EQASs, were evaluated according to individual scheme scoring criteria. The same results were then used to produce z scores using scheme-based between-laboratory SDs as the estimate of variability to determine whether simple performance-derived quality specifications produced better agreement among schemes. Results: The schemes gave conflicting assessments of participants’ performance, and participants judged to be successful by one scheme could be defined as performing inadequately by another. An approach proposed by Kenny et al. (Scand J Clin Lab Invest 1999;59:585), which uses clinical inputs to set targets for analytical imprecision, bias, and total error allowable, was then used to elaborate quality specifications. Conclusions: We suggest that the CLIA ′88 recommendations for blood lead (± 40 μg/L or ± 10% of the target concentration, whichever is the greater) could be used as a quality specification, although a revision to ± 30 μg/L or ± 10% is recommended. For serum aluminum, a suitable quality specification of ± 5 μg/L or ± 20% of the target concentration, whichever is the greater, is suggested. These specifications may be used to compare laboratory performance across schemes.Keywords
This publication has 9 references indexed in Scilit:
- An assessment of contemporary atomic spectroscopic techniques for the determination of lead in blood and urine matricesSpectrochimica Acta Part B: Atomic Spectroscopy, 2001
- Limited seasonality effects on blood lead for a small cohort of female adults and childrenScience of The Total Environment, 2000
- General strategies to set quality specifications for reliability performance characteristicsScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Consensus agreementScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Reference materials and analytical standards to stimulate improved laboratory performance: Experience from the external quality assessment scheme for trace elements in biological samplesMicrochimica Acta, 1996
- Reference materials for measurement of aluminium in biological samplesAnalytical and Bioanalytical Chemistry, 1988
- An external quality assessment scheme for trace elements in biological fluidsJournal of Analytical Atomic Spectrometry, 1986
- Baseline Level of Blood Lead Concentration among Japanese FarmersArchives of Environmental Health: An International Journal, 1985
- Temporal Stability of Blood Lead Concentrations in Adults Exposed Only to Environmental LeadHuman Toxicology, 1984