The Intraocular Lens Tilt and Decentration of Two Different Hydrophobic Aspheric Monofocal Intraocular Lenses
Authors:
- Yasin Sakir GokerYGYasin Sakir Goker
- Serdar OzatesSOSerdar Ozates
- Mustafa KocMKMustafa Koc
- Halil İbrahim AteşoğluHAHalil İbrahim Ateşoğlu
Open Access
- 15 September 2025
- journal article
- Published by Societatea Romana de Oftalmologie in Romanian Journal of Ophthalmology
- Vol. 69 (3) , 357-365
Abstract
To evaluate the clinical performance of two monofocal aspheric intraocular lenses (IOLs) to determine their position in the capsular bag after implantation. This retrospective study collates data from the medical records of patients who underwent cataract surgery and were implanted with two different hydrophobic aspheric monofocal IOLs. A total of 6 visit data were considered in the study, out of which four follow-up visits were on day 1, 7, 30, and 180 postoperatively. The investigational variables included IOL tilt and decentration, uncorrected and best corrected visual acuity (UCVA and BCVA), contrast sensitivity, manifest refraction, intraocular pressure, and posterior capsule opacification. A total of 85 subjects’ medical records were screened. Nineteen subjects discontinued the study due to loss to follow-up. Thirty subjects’ medical records were enrolled in the IOL 1 arm, and 36 subjects’ medical records were enrolled in the IOL 2 arm. There were no statistically significant differences in vertical and horizontal IOL tilt and decentration results between IOL 1 arm and IOL 2 arm. Within-group analysis revealed a significant difference in BCVA at all visits from baseline for both the IOL 1 and IOL 2 arms. In the between-group analysis, there were no significant differences in BCVA at any visit between the IOL 1 and IOL 2 arms. Tilt and decentration of IOL may occur secondary to complicated cataract surgery or following uneventful phacoemulsification. IOL tilt up to 2-3 degrees and decentration of 0.2-0.3 mm are common and not clinically significant. Larger amounts of decentration and tilt deteriorate the optical performance of the IOLs, subsequently affecting the patients’ vision. IOL vertical and horizontal tilt and decentration found in the current study are within the limits reported in the literature. The position of these two lenses in the capsular bag remains very stable at day 180 after implantation.Keywords
This publication has 19 references indexed in Scilit:
- Assessment of tilt and decentration of crystalline lens and intraocular lens relative to the corneal topographic axis using anterior segment optical coherence tomographyPLOS ONE, 2017
- Visual outcomes and optical performance of a monofocal intraocular lens and a new-generation multifocal intraocular lensJournal of Cataract & Refractive Surgery, 2011
- Optical aberrations and alignment of the eye with ageJournal of Vision, 2010
- Comparison of anterior capsule contraction between hydrophobic and hydrophilic intraocular lens modelsGraefe's Archive for Clinical and Experimental Ophthalmology, 2010
- Comparison of the incidence and visual significance of posterior capsule opacification between multifocal spherical, monofocal spherical, and monofocal aspheric intraocular lensesJournal of Cataract & Refractive Surgery, 2009
- Tilt and decentration of spherical and aspheric intraocular lenses: Effect on higher-order aberrationsJournal of Cataract & Refractive Surgery, 2009
- Tilt and decentration of intraocular lenses in vivo from Purkinje and Scheimpflug imagingJournal of Cataract & Refractive Surgery, 2007
- The Thickness of the Aging Human Lens Obtained from Corrected Scheimpflug ImagesOptometry and Vision Science, 2001
- Posterior capsule opacification 3 years after implantation of an AcrySof and a MemoryLens in fellow eyesJournal of Cataract & Refractive Surgery, 2000
- Intraocular lens tilt and decentration: A concern for contemporary IOL designsNepalese Journal of Ophthalmology, 1970