Recognizing, quantifying and managing patient-ventilator asynchrony in invasive and noninvasive ventilation
- 31 May 2018
- journal article
- review article
- Published by Taylor & Francis in Expert Review of Respiratory Medicine
- Vol. 12 (7) , 557-567
Abstract
Introduction: Patient–ventilator asynchrony may occur with modes of partial ventilatory support. Because this problem is associated with worsened outcomes, identifying and managing asynchronies has been recognized as a relevant clinical problem during both invasive and noninvasive (NIV) mechanical ventilation. Areas covered: In this review article, we first describe the different forms of patient–ventilator asynchrony and how they are classified and quantified. Then, we show how these asynchronies can be recognized, considering the techniques used to properly detect asynchronies, by either ventilator waveform observation, or through systems based on more complexes mathematical algorithms, by means of adjunctive signals, such as the electrical activity of the diaphragm or esophageal pressure. Finally, we describe the actions that can be undertaken in order to limit the rate of asynchronies during both invasive ventilation and NIV mechanical ventilation, such as modifications of the ventilator mode and/or settings, variation of the sedation regimen (type and doses), and other technical pitfalls. Expert commentary: Detection of asynchronies is crucial in order to reduce their incidence, adopting adjustments of the ventilator settings, sedation regimen, and other technical pitfalls. It remains to be clarified whether the relationship between high incidence of asynchrony and worsened outcome is causative or just associative.Keywords
This publication has 81 references indexed in Scilit:
- Comparative evaluation of three interfaces for non-invasive ventilation: a randomized cross-over design physiologic study on healthy volunteersCritical Care, 2014
- An automated and standardized neural index to quantify patient-ventilator interactionCritical Care, 2013
- Optimizing Patient-Ventilator Synchrony During Invasive Ventilator Assist in Children and Infants Remains a Difficult Task*Pediatric Critical Care Medicine, 2013
- Impact of Ventilator Adjustment and Sedation–Analgesia Practices on Severe Asynchrony in Patients Ventilated in Assist-Control Mode*Critical Care Medicine, 2013
- Bench comparative evaluation of a new generation and standard helmet for delivering non-invasive ventilationIntensive Care Medicine, 2012
- Polysomnography in stable COPD under non-invasive ventilation to reduce patient–ventilator asynchrony and morning breathlessnessSleep and Breathing, 2011
- Automatic detection of patient-ventilator asynchrony by spectral analysis of airway flowCritical Care, 2011
- On the imperfect synchrony between patient and ventilatorCritical Care, 2011
- Observational study of patient-ventilator asynchrony and relationship to sedation levelJournal of Critical Care, 2009
- Detecting ineffective triggering in the expiratory phase in mechanically ventilated patients based on airway flow and pressure deflection: Feasibility of using a computer algorithm*Critical Care Medicine, 2008