Maintaining the Standard: A Quality Assurance Study for New Equipment in the Michigan ECMO Program
Authors:
- Monika GuptaMGMonika Gupta
- Elaine CooleyECElaine Cooley
- Robert RemenappRRRobert Remenapp
- Robert H. BartlettRBRobert H. Bartlett
- Gail M. AnnichGAGail M. Annich
- 1 September 2007
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 53 (5) , 556-560
Abstract
Activated clotting time (ACT) ranges are targeted to minimize both bleeding and clotting complications during anticoagulation for extracorporeal life support (ECLS) therapy. A change in the equipment used to measure ACT at the bedside prompted a laboratory-based method comparison study in an animal model to compare ACT results from two Hemochron machines. A retrospective, observational cohort study was undertaken to compare bleeding and clotting complications for both "eras" of Hemochron machine. The newer Hemochron Response ACT results were approximately 30 seconds longer than the outgoing Hemochron 401 results at both baseline and heparinization. This prompted a change in the clinical practice standard at the University of Michigan and a shift of the goal ACT range from 180 to 200 seconds to 210 to 230 seconds. Bleeding and clotting complications were less frequent in the Hemochron Response group, although significantly more heparin was infused during the ECLS period. Changes in equipment, which measure important physiologic variables at the bedside, need to be compared with previously established standards to ensure that patient care and safety is maintained, as demonstrated by the longer ACT results obtained with the newer Hemochron Response machine.Keywords
This publication has 10 references indexed in Scilit:
- Comparison of ACT point-of-care measurements: repeatability and agreementPerfusion, 2006
- New technology, old standards: disparate activated clotting time measurements by the Hemochron Jr compared with the standard HemochronThe Annals of Thoracic Surgery, 2004
- Maintaining Adequate Anticoagulation on Extracorporeal Membrane Oxygenation Therapy: Hemochron Junior Low Range versus Hemochron 400The Journal of ExtraCorporeal Technology, 2003
- Low-dose versus high-dose heparinization during arteriovenous carbon dioxide removalPerfusion, 2001
- Hemodynamic effect of a low-resistance artificial lung in series with the native lungs of sheepThe Annals of Thoracic Surgery, 2000
- Measuring agreement in method comparison studiesStatistical Methods in Medical Research, 1999
- Total Respiratory Support With Tidal Flow Extracorporeal Circulation in Adult SheepASAIO Journal, 1997
- HEMODYNAMIC EFFECTS AND CARDIAC FUNCTION DURING TIDAL FLOW EXTRACORPOREAL CIRCULATION (TF ECC)ASAIO Journal, 1997
- Effects of moderate hypothermia on O2 consumption at various O2 deliveries in a sheep modelJournal of Applied Physiology, 1992
- Prolonged Partial Venoarterial Bypass:Physiologic, Biochemical, and Hematologic ResponsesAnnals of Surgery, 1974