Abstract
Commercial macroscopic test-strips (dipsticks) that indirectly detect urine leukocytes by quantifying leukocyte esterase (LE) activity have been advocated as a simple, rapid, and inexpensive alternative to microscopic examination for detection of significant pyuria in urine specimens. The purpose of this study was to evaluate the diagnostic performance of a commercial LE test-strip for detection of feline pyuria. Two hundred and thirteen consecutive urine specimens were collected from 188 different feline patients and analyzed for LE activity with a LE test-strip (Multistix 2 Reagent Strips:Ames Division, Bayer Corp., Elkhart, IN). Results of the LE test-strip were compared with those of standard urine biochemical and microscopic sediment evaluations. Compared with urine sediment leukocyte counts, the LE test-strip had a sensitivity of 77%, a specificity of 34%, positive and negative predictive values of 14 and 91% respectively and an overall test efficiency of 39%. Multivariable logistic regression analysis did not reveal significant associations between pyuria (>5 WBC/hpf) and a positive LE test-strip reaction; however, hematuria, lipiduria, increasing age, and decreasing urine specific gravity were associated with a significantly increased risk for positive LE test-strip reactions. We conclude that the LE test-strip evaluated in this study is highly nonspecific for defection of significant pyuria in feline urine specimens and should not replace routine microscopic urine sediment examination in this species.Keywords
This publication has 13 references indexed in Scilit:
- The development and structure of mature neutrophils1994
- Potent leukocidal action of Escherichia coli hemolysin mediated by permeabilization of target cell membranes.Journal of Experimental Medicine, 1989
- The Urinalysis: A Critical AppraisalMedical Clinics of North America, 1987
- The Detection of Leukocyte Esterase Activity in Urine with a New Reagent StripAmerican Journal of Clinical Pathology, 1987
- Clinical Significance of Esterases in ManClinical Pharmacokinetics, 1985
- New and rapid method for detection of pyuria by leukocyte esterase reactionUrology, 1984
- Purification and Characterization of Leucocidin from Pseudomonas aeruginosaJournal of General Microbiology, 1976
- PYURIA: DEPOSIT IN HIGH‐POWER MICROSCOPIC FIELD—WBC/HPF—VERSUS WBC/mm3 IN COUNTING CHAMBER.Acta Medica Scandinavica, 1973
- Survival of urinary leucocytesJournal of Clinical Pathology, 1966
- STAPHYLOCOCCAL LEUKOCIDINAnnals of the New York Academy of Sciences, 1965