Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients
- 20 November 2001
- journal article
- other
- Published by Wolters Kluwer Health in Circulation
- Vol. 104 (21) , 2533-2538
Abstract
Background Defibrillation thresholds (DFT) with standard implantable cardioverter-defibrillator leads in the right ventricle (RV) may be determined by weak shock field intensity in the myocardium of the left ventricle (LV). Adding a shocking electrode in a coronary vein on the middle of the LV free wall, thereby establishing biventricular defibrillation, substantially reduced defibrillation requirements in animals. We investigated the feasibility of this approach in 24 patients receiving an implantable cardioverter-defibrillator using a prototype over-the-wire temporary LV defibrillation lead. Methods and Results The LV lead was inserted through the coronary sinus, using a guide catheter and guidewire, into a posterior or lateral coronary vein whose location was determined by retrograde venography. Paired DFT testing compared a standard system (RV to superior vena cava plus can emulator [SVC+Can], 60% tilt biphasic shock) to a system including the LV lead. The biventricular system was tested with a dual-shock waveform (20% tilt monophasic shock from LV→SVC+Can, then 60% tilt biphasic shock from RV→SVC+Can). Twenty patients completed DFT testing. Venography and LV lead insertion time was 46±40 minutes. The biventricular system reduced mean DFT by 45% (8.9±1.1 J versus 4.9±0.5 J, P Conclusions Internal defibrillation using a transvenously inserted LV lead is feasible, produces significantly lower DFTs, and seems safe under the conditions tested. Biventricular defibrillation may be a useful option for reducing DFTs or could be added to an LV pacing lead for heart failure.Keywords
This publication has 16 references indexed in Scilit:
- Human Experience with Transvenous Biventricular Defibrillation Using an Electrode in a Left Ventricular VeinPacing and Clinical Electrophysiology, 2002
- Transvenous left ventricular lead implantation with the EASYTRAK lead system: the European experienceThe American Journal of Cardiology, 2000
- Marked reduction of ventricular defibrillation threshold by application of an auxiliary shock to a catheter electrode in the left posterior coronary vein of dogs.Journal of Cardiovascular Electrophysiology, 2000
- Transvenous biventricular pacing for heart failure: can the obstacles be overcome?The American Journal of Cardiology, 1999
- Defibrillation Efficacy of Different Electrode Placements in a Human Thorax ModelPacing and Clinical Electrophysiology, 1999
- The Subcutaneous Array: A New Lead Adjunct for the Transvenous ICD to Lower Defibrillation ThresholdsPacing and Clinical Electrophysiology, 1995
- Combination biphasic waveform plus sequential pulse defibrillation improves defibrillation efficacy of a nonthoracotomy lead systemJACC, 1994
- Effective Defibrillation in Pigs Using Interleaved and Common Phase Sequential Biphasic ShocksPacing and Clinical Electrophysiology, 1993
- Defibrillation Threshold: Clinical Utility and Therapeutic ImplicationsPacing and Clinical Electrophysiology, 1992
- Sequential Pulse Countershock Between Two Transvenous Catheters: Feasibility, Safety, and EfficacyPacing and Clinical Electrophysiology, 1989