Abstract
This report describes an 85 year‐old man who underwent percutaneous aortic valve replacement (PAVR). With a logistic euroSCORE of 37%, the patient had been refused surgical aortic valve replacement because of an unacceptably high peri‐operative risk. During the PAVR procedure, severe resistance was encountered when advancing the 21 Fr delivery catheter through the left iliac artery despite pre‐dilatation with a 7 mm balloon. Following this, PAVR was promptly achieved without difficulty, with excellent valve positioning, no peri‐valvular leak and good hemodynamics. However, transesophageal echocardiography revealed a mobile echogenic mass within the outflow tract of the left ventricle. The mass was retrieved with a cardiac bioptome manipulated via the left femoral artery through a 9 Fr sheath. A right cerebral ischemic stroke manifested shortly after the post‐procedure and the patient died on the fourth post‐operative day. Post‐mortem findings revealed a left subclavian artery occlusion by iliac vascular tissue. This report highlights the imperative for device‐specific vascular access screening criteria and the need to minimize device size in order to safely accomplish PAVR.