Abstract
Purpose:The purpose was to evaluate individually shaped contrast media (CM) delivery in CT pulmonary angiography (CTPA) for suspected pulmonary embolism.Materials and Methods:100 consecutive emergency patients with clinical suspicion for pulmonary embolism were evaluated. High-pitch CTPA was performed on a 2nd generation dual-source CT using the following parameters: 100kV, 200-250mAsref, rot.time 0.28s, 128x0.6mm col., image reconstruction 1.0/0.8mm (B30f). Group 1 (n=50) then received a fixed CM bolus (300mgI/ml, volume=90ml, flow rate=6ml/s); group 2 (n=50) received a body weight adapted CM bolus determined by dedicated contrast injection software. For analysis, groups were further subdivided into low (40-75kg) and high (76-117kg) weight groups. Technical image quality was graded using a four-point Likert scale (1=non-diagnostic; 2=diagnostic; 3=good and 4=excellent image quality) at the level of the pulmonary trunk and pulmonary arteries. Objective image quality analysis was done measuring contrast en...