An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose
Open Access
- 9 September 2019
- journal article
- research article
- Published by Wiley in Journal of Diabetes Research
- Vol. 2019, 1-7
Abstract
Aims. To evaluate the performance of a factory-calibrated flash glucose monitoring system in children with diabetes compared to venous blood glucose (BG). Methods. A total of 13 hospitalized participants newly diagnosed with type 1 diabetes, aged 1~14 years old, were involved in the study. Sensor glucose measurements on days 2, 3, 6, 7, 12, and 13 of wear were compared with venous BG. During these days, the venous BG results were obtained either 4 or 7 times per day. Results. The accuracy was evaluated against venous BG, with 469 of 469 (100.0%) sensor and venous BG pairs within consensus error grid zones A and B, including 94.7% in zone A. The overall mean absolute relative difference (MARD) was 11.67%. The MARD of blood glucose lower than 4.0 mmol/L () was higher than blood glucose between 4 and 10 mmol/L () and blood glucose higher than 10 mmol/L (). Compared to venous BG, the MARDs of wear days 2, 3, 6, 7, 12, and 13 were 11.53%, 9.66%, 11.79%, 10.89%, 13.18%, and 13.92%, respectively, with no statistically significant difference (). The median ARD was highest when the glucose decreased >0.11 mmol/L/min (20.27%) and lower than 10.00% when the glucose changed between 0.06 and 0.11 mmol/L/min, changed 0.11 mmol/L/min. Conclusions. The accuracy of the system is good and remains stable over 14 days of wear; however, the accuracy depends on the glucose level and rates of glucose concentration changes.
Keywords
Funding Information
- Beijing Municipal Administration of Hospitals (PX2016009, ZYLX201821, BCHYIPA-2016-06, 2016YFC1305304)
- Beijing Municipal Administration of Hospitals Clinical Medicine Development of Special Funding Support (PX2016009, ZYLX201821, BCHYIPA-2016-06, 2016YFC1305304)
- Beijing Children’s Hospital Young Investigator Program (PX2016009, ZYLX201821, BCHYIPA-2016-06, 2016YFC1305304)
- National Key Research and Development Program of China (PX2016009, ZYLX201821, BCHYIPA-2016-06, 2016YFC1305304)
This publication has 14 references indexed in Scilit:
- An alternative sensor-based method for glucose monitoring in children and young people with diabetesArchives of Disease in Childhood, 2017
- IntroductionDiabetes Care, 2016
- A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With DiabetesJournal of Diabetes Science and Technology, 2016
- The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring SystemDiabetes Technology & Therapeutics, 2015
- A New Horizon for Glucose MonitoringHormone Research in Paediatrics, 2015
- Use of Subcutaneous Interstitial Fluid Glucose to Estimate Blood Glucose: Revisiting Delay and Sensor OffsetJournal of Diabetes Science and Technology, 2010
- A Tale of Two Compartments: Interstitial Versus Blood Glucose MonitoringDiabetes Technology & Therapeutics, 2009
- Timing of Changes in Interstitial and Venous Blood Glucose Measured With a Continuous Subcutaneous Glucose SensorDiabetes, 2003
- Continuous Glucose Monitoring with Glucose Sensors: Calibration and Assessment CriteriaDiabetes Technology & Therapeutics, 2003
- Does fall in tissue glucose precede fall in blood glucose?Diabetologia, 1996