Reliability Analysis of 16-MDCT in Preoperative Evaluation of Total Knee Arthroplasty and Comparison With Intraoperative Measurements

Authors:

Abstract
OBJECTIVE. The aim of our study was to determine how consistent measurements of the distal femoral condyle on 16-MDCT are with intraoperative measurements and the reliability of 16-MDCT for the preoperative planning of total knee arthroplasty.MATERIALS AND METHODS. Between August 2003 and March 2004, 33 consecutive patients (two men and 31 women; age range, 53-89 years; mean age, 71 years) presenting with osteoarthritis underwent 16-MDCT (Mx 8000 IDT) of a total of 53 knees before total knee arthroplasty. The prospective analysis included measurements of transepicondylar distance, maximum anteroposterior dimension of medial and lateral femoral condyles, and trochlear width on a Rapidia workstation. To increase reliability, we repeated the measurements on a CT workstation after 2 months and compared them with the previous values. The values measured on the CT workstation were compared with the measurements obtained in the intraoperative field. Statistical analysis was performed using kappa statistics. A p ...