A method to correct background phase offset for phase‐contrast MRI in the presence of steady flow and spatial wrap‐around artifact
Authors:
- 15 November 2018
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 81 (4) , 2424-2438
Abstract
Purpose: Background phase offsets in phase‐contrast MRI are often corrected using polynomial regression; however, correction performance degrades when temporally invariant outliers such as steady flow or spatial wrap‐around artifact are present. We describe and validate an iterative method called automatic rejection of temporally invariant outliers (ARTO), which excludes these outliers from the fitting process.Methods: The ARTO method iteratively removes pixels with large polynomial regression errors analyzed by a Gaussian mixture model fitting of the residual distribution. A total of 150 trials of a simulated phantom (75 with wrap‐around artifact) and 125 phase‐contrast MRI cines from 22 healthy subjects (48 with wrap‐around artifact) were used for validation. Background phase offsets were corrected using second‐order weighted regularized least squares (WRLS) with and without ARTO. Flow volumes after WRLS and WRLS+ARTO corrections were compared with the known truth (phantom) and stationary phantom reference (in vivo) using Bland‐Altman analysis. The ratio between the pulmonary flow and the systemic flow was also computed in a subset of 6 subjects.Results: In the simulated phantom, compared with WRLS and no correction, correction with WRLS+ARTO produced superior agreement in volumetric flow quantification with the known truth. In vivo, WRLS+ARTO also produced superior agreement with stationary phantom‐corrected volumetric flow compared with WRLS and no correction. In data sets with wrap‐around artifact, WRLS produced significantly larger variance in the pulmonary flow and systemic flow ratio than stationary phantom correction (P= .0008).Conclusion: The proposed method provides automatic exclusion of temporally invariant outliers and produces flow quantification results comparable to stationary phantom correction.Keywords
Funding Information
- National Institutes of Health (R21EB021655, R21EB022277)
This publication has 26 references indexed in Scilit:
- Accuracy of four-dimensional phase-contrast velocity mapping for blood flow visualizations: a phantom studyActa Radiologica, 2013
- A multi-center inter-manufacturer study of the temporal stability of phase-contrast velocity mapping background offset errorsJournal of Cardiovascular Magnetic Resonance, 2012
- Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurementsJournal of Cardiovascular Magnetic Resonance, 2010
- Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonanceJournal of Cardiovascular Magnetic Resonance, 2010
- Design and validation of Segment - freely available software for cardiovascular image analysisBMC Medical Imaging, 2010
- Improved accuracy in flow mapping of congenital heart disease using stationary phantom techniqueJournal of Cardiovascular Magnetic Resonance, 2009
- Cardiology Is FlowCirculation, 2006
- Efficient foldover suppression using SENSEMagnetic Resonance Materials in Physics, Biology and Medicine, 2004
- Semiautomated method for noise reduction and background phase error correction in MR phase velocity dataJournal of Magnetic Resonance Imaging, 1993
- Flow-velocity imaging from linear regression of phase images with techniques for reducing eddy-current effectsJournal of Magnetic Resonance (1969), 1990