Commissioning of a gantry beamline with rotator at a synchrotron-based ion therapy center
Open Access
- 26 February 2024
- journal article
- research article
- Published by American Physical Society (APS) in Physical Review Accelerators and Beams
- Vol. 27 (2) , 023503
Abstract
This paper provides an overview of the worldwide first commissioning of a gantry beamline with a rotator at the MedAustron synchrotron-based proton/ion cancer therapy facility in Wiener Neustadt, Austria. The gantry beamline consists of the high energy beam transfer (HEBT) line and the gantry beam transport system. It transports the beam from the synchrotron to the gantry-room isocenter. The HEBT transports the beam from the synchrotron to the gantry entrance, which is the coupling point between the HEBT and the gantry. The rotator is one of the HEBT modules, thus it is an integral part of the gantry beamline. The MedAustron rotator is the worldwide first rotator system used to match slowly extracted asymmetric beams from the synchrotron to the rotating gantry. In this paper, main attention is paid to ion-optical and beam-alignment aspects of the beamline commissioning. A novel orbit-correction and beam-alignment technique has been developed specifically for the beamline with the rotator. While the theoretical concept of the rotator has existed for almost two decades, the MedAustron rotator is the first hardware implementation of this concept all over the world. The presented overview of the beamline commissioning includes a description of the principal technical solutions and main results of the first beam-transport measurements. Since the measured beam size and beam position agree well with theoretical predictions, one can conclude that the proof-of-concept of the rotator-matching has been successfully accomplished. Published by the American Physical Society 2024Keywords
Funding Information
- Vedecká Grantová Agentúra MŠVVaŠ SR a SAV (VEGA 1/0130/20)
- European Regional Development Fund (ITMS2014+: 313011BUH7)
This publication has 26 references indexed in Scilit:
- Recent progress of a superconducting rotating-gantry for carbon-ion radiotherapyNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2017
- Beam commissioning of a superconducting rotating-gantry for carbon-ion radiotherapyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Current and future accelerator technologies for charged particle therapyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Overview of the FLUKA codeAnnals of Nuclear Energy, 2015
- The National Centre for Oncological Hadrontherapy (CNAO): Status and perspectivesPhysica Medica, 2015
- Development of a superconducting rotating-gantry for heavy-ion therapyNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2013
- The MedAustron projectNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2007
- Compensation of the asymmetric phase–space distribution for a slowly extracted beam from a synchrotronNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Beam-transport study of an isocentric rotating ion gantry with minimum number of quadrupolesNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2005
- Optics design of the extraction lines for the MedAustron hadron therapy centreNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2005