Review of Reported Clinical Information System Adverse Events in US Food and Drug Administration Databases
Authors:
- S.L. JonesSJS.L. Jones
- D.F. SittigD.F. Sittig
- 1 January 2011
- journal article
- review article
- Published by Georg Thieme Verlag KG in Applied Clinical Informatics
- Vol. 02 (01) , 63-74
Abstract
Summary: Background: The US FDA has been collecting information on medical devices involved in significant adverse advents since 1984. These reports have been used by researchers to advise clinicians on potential risks and complications of using these devices. Objective: Research adverse events related to the use of Clinical Information Systems (CIS) as reported in FDA databases. Methods: Three large, national, adverse event medical device databases were examined for reports pertaining to CIS. Results: One hundred and twenty unique reports (from over 1.4 million reports) were found, representing 32 manufacturers. The manifestations of these adverse events included: missing or incorrect data, data displayed for the wrong patient, chaos during system downtime and system unavailable for use. Analysis of these reports illustrated events associated with system design, implementation, use, and support. Conclusion: The identified causes can be used by manufacturers to improve their products and by clinical facilities and providers to adjust their workflow and implementation processes appropriately. The small number of reports found indicates a need to raise awareness regarding publicly available tools for documenting problems with CIS and for additional reporting and dialog between manufacturers, organizations, and users.Keywords
This publication has 14 references indexed in Scilit:
- Commentary on “secondary use of electronic health record data: Spontaneous triggered adverse drug event reporting” by linder et al.Pharmacoepidemiology and Drug Safety, 2010
- Secondary use of electronic health record data: spontaneous triggered adverse drug event reportingPharmacoepidemiology and Drug Safety, 2010
- Monitoring and Evaluating the Use of Electronic Health Records2010
- Lessons Premier Hospitals Learned About Implementing Electronic Health RecordsHealth Affairs, 2010
- Safe Electronic Health Record Use Requires a Comprehensive Monitoring and Evaluation FrameworkJAMA, 2010
- Eight Rights of Safe Electronic Health Record UseJAMA, 2009
- Types of Unintended Consequences Related to Computerized Provider Order EntryJournal of the American Medical Informatics Association, 2006
- Computerized Provider Order Entry Implementation: No Association With Increased Mortality Rates in an Intensive Care UnitPediatrics, 2006
- Unexpected Increased Mortality After Implementation of a Commercially Sold Computerized Physician Order Entry SystemPediatrics, 2005
- Recommendations for Responsible Monitoring and Regulation of Clinical Software SystemsJournal of the American Medical Informatics Association, 1997