Model predictive control of a robotically actuated delivery sheath for beating heart compensation
Open Access
- 1 February 2017
- journal article
- research article
- Published by SAGE Publications in The International Journal of Robotics Research
- Vol. 36 (2) , 193-209
Abstract
Minimally invasive surgery (MIS) during cardiovascular interventions reduces trauma and enables the treatment of high-risk patients who were initially denied surgery. However, restricted access, reduced visibility and control of the instrument at the treatment locations limits the performance and capabilities of such interventions during MIS. Therefore, the demand for technology such as steerable sheaths or catheters that assist the clinician during the procedure is increasing. In this study, we present and evaluate a robotically actuated delivery sheath (RADS) capable of autonomously and accurately compensating for beating heart motions by using a model-predictive control (MPC) strategy. We develop kinematic models and present online ultrasound segmentation of the RADS that are integrated with the MPC strategy. As a case study, we use pre-operative ultrasound images from a patient to extract motion profiles of the aortic heart valve (AHV). This allows the MPC strategy to anticipate for AHV motions. Further, mechanical hysteresis in the steering mechanism is compensated for in order to improve tip positioning accuracy. The novel integrated system is capable of controlling the articulating tip of the RADS to assist the clinician during cardiovascular surgery. Experiments demonstrate that the RADS follows the AHV motion with a mean positioning error of 1.68 mm. The presented modelling, imaging and control framework could be adapted and applied to a range of continuum-style robots and catheters for various cardiovascular interventions.Keywords
Funding Information
- Stichting voor de Technische Wetenschappen (iMIT-MULTI)
This publication has 43 references indexed in Scilit:
- Image-based hysteresis reduction for the control of flexible endoscopic instrumentsMechatronics, 2013
- Transapical Versus Transfemoral Aortic Valve Implantation: A Comparison of Survival and SafetyThe Annals of Thoracic Surgery, 2011
- Transapical Beating Heart Mitral Valve RepairCirculation: Cardiovascular Interventions, 2010
- Technical considerations to avoid pitfalls during transapical aortic valve implantationThe Journal of Thoracic and Cardiovascular Surgery, 2010
- Steerable Laparoscopic Cable-Ring ForcepsJournal of Medical Devices, 2010
- Percutaneous mitral valve repair with the MitraClip system: acute results from a real world settingEuropean Heart Journal, 2010
- GPU based real-time instrument tracking with three-dimensional ultrasoundMedical Image Analysis, 2007
- Basic physics of ultrasound imagingCritical Care Medicine, 2007
- Manual control and tracking—a human factor analysis relevant for beating heart surgeryThe Annals of Thoracic Surgery, 2002
- Adaptive generalized predictive controller for MIMO systemsInternational Journal of Control, 1989