Wall thickness of the pulmonary vein-left atrial junction rather than electrical information as the major determinant of dormant conduction after contact force-guided pulmonary vein isolation
Authors:
- 25 May 2016
- journal article
- clinical trial
- Published by Springer Nature in Journal of Interventional Cardiac Electrophysiology
- Vol. 46 (3) , 325-333
Abstract
The usefulness of electrogram (EGM)-based information has been reported for assessing lesion transmurality during atrial fibrillation (AF) ablation, but the wall thickness of the pulmonary vein-left atrial (PV-LA) junction has not been considered. We conducted a study to evaluate the relation between PV-LA junction wall thickness and the presence of adenosine triphosphate (ATP)-provoked dormant PV conduction. Eighteen AF patients underwent extensive encircling pulmonary vein isolation (EEPVI) with a target CF of >10 g. RF energy was delivered point-by-point at a power setting of 25–30 W for 30 s, and EGM-based information (change in filtered unipolar EGM morphology and bipolar EGM amplitude), decrease in impedance, CF, and CT-based measurement of the PV-LA junction wall thickness were characterized at sites of ATP-provoked dormant conduction. After EEPVI, ATP-induced dormant conduction was observed at 12 of the 288 PV sites (8 segments per ipsilateral PVs × 2 × 18 patients). Of the 974 ablation points, 72 were located at dormant conduction sites and were strongly associated with thickened PV-LA junction walls (1.02± 0.23 vs. 0.86 ± 0.26 mm, p < 0.0001) and decreased impedance (13.3 ± 6.4 vs. 14.9 ± 7.1 Ω, p = 0.0498) but not with EGM-based information or CF. Multivariate analysis identified the thickened PV-LA junction wall as the strongest predictor of dormant conduction. A thickened PV-LA junction wall is a robust predictor of ATP-provoked dormant conduction; EGM-based information appears to be insufficient for ensuring adequate lesions during CF-guided EEPVI.Keywords
This publication has 26 references indexed in Scilit:
- Adenosine-guided pulmonary vein isolation for the treatment of paroxysmal atrial fibrillation: an international, multicentre, randomised superiority trialThe Lancet, 2015
- Pulmonary vein isolation using “contact force” ablation: The effect on dormant conduction and long-term freedom from recurrent atrial fibrillation—A prospective studyHeart Rhythm, 2014
- Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: A prospective case control studyInternational Journal of Cardiology, 2013
- The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA studyHeart Rhythm, 2012
- Local unipolar and bipolar electrogram criteria for evaluating the transmurality of atrial ablation lesions at different catheter orientations relative to the endocardial surfaceHeart Rhythm, 2010
- ATP‐Induced Dormant Pulmonary Veins Originating from the Carina Region After Circumferential Pulmonary Vein Isolation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2010
- Clinical Implications of Reconnection Between the Left Atrium and Isolated Pulmonary Veins Provoked by Adenosine Triphosphate after Extensive Encircling Pulmonary Vein IsolationJournal of Cardiovascular Electrophysiology, 2007
- Recovered Pulmonary Vein Conduction as a Dominant Factor for Recurrent Atrial Tachyarrhythmias After Complete Circular Isolation of the Pulmonary VeinsCirculation, 2005
- “Dormant” Pulmonary Vein Conduction Revealed by Adenosine after Ostial Radiofrequency Catheter AblationJournal of Cardiovascular Electrophysiology, 2004
- Monitoring the Local Electrogram at the Ablation Site during Radiofrequency Application for Common Atrial FlutterJapanese Circulation Journal, 1998