Relationship between stimulus amplitude, stimulus frequency and neural damage during electrical stimulation of sciatic nerve of cat
Authors:
- 1 May 1995
- journal article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 33 (3) , 426-429
Abstract
The relation is investigated between stimulus frequency, stimulus pulse amplitude and the neural damage induced by continuous stimulation of the cat's sciatic nerve. The chronically implanted electrodes were pulsed continuously and the effects of the electrical stimulation were quantified as the amount of early axonal degeneration (EAD) present in the nerves seven days after the continuous stimulation. The primary effect of stimulating at 100 Hz rather than 50 Hz was to cause an increase in the slope of the plot of the amount of EAD versus stimulus lower. There was a small amount of EAD in three of the nerves stimulated at 20 Hz, but there was no detectable correlation between the amount of EAD and the stimulus amplitude. This suggests that continuous electrical stimulation of peripheral nerves at a low frequency induce little or no neural damage, even if the stimulus amplitude is very high. A preliminary presentation of the results has been made elsewhere (Agnew et al., 1993)Keywords
This publication has 4 references indexed in Scilit:
- Damage in peripheral nerve from continuous electrical stimulation: Comparison of two stimulus waveformsMedical & Biological Engineering & Computing, 1992
- Histologic and physiologic evaluation of electrically stimulated peripheral nerve: Considerations for the selection of parametersAnnals of Biomedical Engineering, 1989
- The Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve StimulationIEEE Transactions on Biomedical Engineering, 1983
- The response of the myelinated nerve fiber to short duration biphasic stimulating currentsAnnals of Biomedical Engineering, 1979