Accurate measurement of the dynamic response of a scanning electronic portal imaging device
Authors:
- Monique H. P. SmitsmansMSMonique H. P. Smitsmans
- Kenneth G. A. GilhuijsKGKenneth G. A. Gilhuijs
- Marcel van HerkMHMarcel van Herk
- 9 March 2001
- journal article
- Published by Wiley in Medical Physics
- Vol. 28 (3) , 310-316
Abstract
An important condition for the safe introduction of dynamic intensity modulated radiotherapy (IMRT) using a multileaf collimator (MLC) is the ability to verify the leaf trajectories. In order to verify IMRT using an electronic portal imaging device (EPID), the EPID response should be accurate and fast. Noninstantaneous dynamic response causes motion blurring. The aim of this study is to develop a measurement method to determine the magnitude of the geometrical error as a result of motion blurring for imagers with scanning readout. The response of a liquid-filled ionization chamber EPID, as an example of a scanning imager, on a moving beam is compared with the response of a diode placed at the surface of the EPID. The signals are compared under the assumption that all EPID rows measure the same dose rate when a straight moving field edge is imaged. The measurements are performed at several levels of attenuation to investigate the influence of dose rate on the response of the detector. The accuracy of the measurement method is better than 0.25 mm. We found that the liquid-filled ionization chamber EPID does not suffer from significant motion blurring under clinical circumstances. Using a maximum gradient edge detector to determine the field edge in an image obtained by a liquid-filled ionization chamber EPID, errors smaller than 1 mm are found at a dose rate of 105 MU/min and a field edge speed of 1.1 cm/s. The errors reduce at higher dose rates. The presented method is capable of quantifying the geometrical errors in determining the position of the edge of a moving field with subpixel accuracy. The errors in field edge position determined by a liquid-filled ionization chamber EPID are negligible in clinical practice. Consequently, these EPIDs are suitable for geometric IMRT verification, as far as dynamic response is concerned.Keywords
This publication has 16 references indexed in Scilit:
- Leaf position verification during dynamic beam delivery: A comparison of three applications using electronic portal imagingMedical Physics, 2000
- Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging deviceMedical Physics, 1999
- Independent verification using portal imaging of intensity‐modulated beam delivery by the dynamic MLC techniqueMedical Physics, 1998
- Quality assurance for dynamic multileaf collimator modulated fields using a fast beam imaging systemMedical Physics, 1997
- Testing of dynamic multileaf collimationMedical Physics, 1996
- Dosimetric verification of intensity‐modulated fieldsMedical Physics, 1996
- Dynamic X-ray compensation for conformal radiotherapy by means of multi-leaf collimationRadiotherapy and Oncology, 1994
- An analytical solution for the dynamic control of multileaf collimatorsPhysics in Medicine & Biology, 1994
- The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimationPhysics in Medicine & Biology, 1992
- Radiation field edge detection in portal imagesPhysics in Medicine & Biology, 1991