Comparison of different methods to assess alveolar cleft defects in cone beam CT images

Authors:

Abstract
Objectives:This study aimed to evaluate the accuracy of three different methods for assessing the volume of cleft defects in CBCT images. The influence of field of view (FOV) and voxel sizes was also assessed.Methods:Using three radiopaque plastic skulls, unilateral defects were created to mimic alveolar clefts and were filled with wax following the contralateral side contours. They were scanned in a CBCT unit using four different acquisition protocols, varying FOV and voxel sizes. Using three different methods, the defect/wax volume was evaluated on the images by defining: (1) width, height and facial-palatal length of the defect in maximum intensity projection; (2) the areas of the defect on axial slices; (3) the threshold and segmentation of the region of interest. The values obtained from each method using different acquisition protocols were compared to the real volume of the wax (gold standard) using ANOVA and Tukey’s test.Results:Methods 2 and 3 did not differ from the gold standard (p>0.05). Conve...

This publication has 23 references indexed in Scilit: