Thermal drift reduction with multiple bias current for MOSFET dosimeters
- 23 May 2011
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (12) , 3535-3550
Abstract
New thermal compensation methods suitable for p-channel MOSFET (pMOS) dosimeters with the usual dose readout procedure based on a constant drain current are presented. Measuring the source-drain voltage shifts for two or three different drain currents and knowing the value of the zero-temperature coefficient drain current, I(ZTC), the thermal drift of source-drain or threshold voltages can be significantly reduced. Analytical expressions for the thermal compensation have been theoretically deduced on the basis of a linear dependence on temperature of the parameters involved. The proposed thermal modelling has been experimentally proven. These methods have been applied to a group of ten commercial pMOS transistors (3N163). The thermal coefficients of the source-drain voltage and the threshold voltage were reduced from -3.0 mV °C(-1), in the worst case, down to -70 µV °C(-1). This means a thermal drift of -2.4 mGy °C(-1) for the dosimeter. When analysing the thermal drifts of all the studied transistors, in the temperature range from 19 to 36 °C, uncertainty was obtained in the threshold voltage due to a thermal drift of ±9 mGy (2 SD), a commonly acceptable value in most radiotherapy treatments. The procedures described herein provide thermal drift reduction comparable to that of other technological or numerical strategies, but can be used in a very simple and low-cost dosimetry sensor.Keywords
This publication has 28 references indexed in Scilit:
- Readout techniques for linearity and resolution improvements in MOSFET dosimetersSensors and Actuators A: Physical, 2010
- Monte Carlo simulation using the PENELOPE code with an ant colony algorithm to study MOSFET detectorsPhysics in Medicine & Biology, 2009
- Evaluation of a low-cost commercial mosfet as radiation dosimeterSensors and Actuators A: Physical, 2006
- Clinical application of the OneDose™ Patient Dosimetry System for total body irradiationPhysics in Medicine & Biology, 2005
- Radiation-induced statistical uncertainty in the threshold voltage measurement of MOSFET dosimetersPhysics in Medicine & Biology, 2004
- Effects of temperature variation on MOSFET dosimetryPhysics in Medicine & Biology, 2004
- Clinical dosimetry with MOSFET dosimeters to determine the dose along the field junction in a split beam techniqueRadiotherapy and Oncology, 2003
- Neutron detection at the extremes of sensitivity in the cosmic environmentIEEE Transactions on Nuclear Science, 2000
- Strategies for MilliRad sensitivity in PMOS dosimetersIEEE Transactions on Nuclear Science, 1998
- On-chip p-MOSFET dosimetry (CMOS ICs)IEEE Transactions on Nuclear Science, 1993