Determination of Al, Cr, Fe, Zn, Cd, Pb and Bi in Crude Drugs by Inductively Coupled Plasma Atomic Emission Spectrometry after Coprecipitation with Yttrium Phosphate

Authors:

Abstract
An ovel method that combines inductively cou- pled plasma atomic emission spectrometry (ICP- AES) with a coprecipitation technique to determine trace and sub-trace elements in crude drugs is de- scribed herein. Yttrium phosphate quantitatively coprecipitated seven elements (Al, Cr, Fe, Zn, Cd, Pb and Bi) at pH 6 in a solution prepared from microwave-digested samples of crude drugs. Some matrix elements such as Na, K, Mg and Ca were effectively removed by this process. The thus co- precipitated elements were then determined by ICP- AES with yttrium as an internal standard element. The detection limits (3σ, n = 10) were 0.17mg/kg for Al, 0.004mg/k gf or Cr, 0.03mg/kg for Fe, 0.13mg/kg for Zn, 0.004mg/kg for Cd, 0.19mg/kg for Pb and 0.06mg/kg for Bi. The proposed method was success- fully utilized to determine the concentration of the above-mentioned seven elements in standard refer- ence materials (National Institute of Standards Tech- nology (NIST) SRM1515, SRM1547 and SRM1575a) and seven kinds of crude drugs.