Abstract
Purpose: The purposes of this study were to 1) measure the effects of Kalman-weighted averaging methods on ABR threshold, latency and amplitude; 2) translate lab findings to the clinical setting; and 3) estimate cost savings when ABRs can be obtained in non-sedated infants. Method: ABRs were recorded in 40 adults with normal hearing during induced motor noise conditions using the Kalman-weighted averaging method implemented on a commercial system, the Vivosonic Integrity™. The device was then used to test 34 infants in awake and asleep states. The advantages of the Kalman-weighted averaging method were modeled in terms of time saved for conducting an ABR evaluation. Results: Kalman-weighted ABR threshold estimates were 6-7 dB lower than with conventional methods during induced motor noise. When used to obtain ABRs in infants who were awake, the number of sweeps required to obtain a result was significantly greater than required for a sleeping infant, but well within the range for clinical application. Conclusions: The use of Kalman-weighted averaging provides a measurable advantage over conventional methods and may reduce costs for the pediatric audiology practice.

This publication has 17 references indexed in Scilit: