Image registration based 3D TEE-EM calibration
Authors:
- 1 April 2010
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 19457928,p. 1209-1212
Abstract
Real-time 3D trans-esophageal echo (TEE) is being integrated into routine clinical practice for minimally invasive cardiac therapy. Electromagnetically (EM) tracking of the TEE probe inside the body may facilitate intraprocedural heart surgery and therapy on a beating heart. This requires an accurate calibration between the 3D TEE and the EM tracking system. The objective of calibration is to find the transformation that converts the coordinates of voxels in ultrasound volumes into the coordinate system of a position sensor attached to the TEE probe. An image registration based ultrasound calibration method is presented and validated using phantoms. The experiments indicate a post-calibration image reconstruction precision of 1.56 mm and accuracy of 0.85 mm (relative accuracy of 1.86%), and TRE of 2.37 mm. The potential impact of this work is that it can provide an easy to use calibration for clinical use.Keywords
This publication has 6 references indexed in Scilit:
- A novel closed form solution for ultrasound calibration2005
- A review of calibration techniques for freehand 3-D ultrasound systemsUltrasound in Medicine & Biology, 2005
- A screw motion approach to uniqueness analysis of head-eye geometry2002
- Hand-Eye Calibration Using Dual QuaternionsThe International Journal of Robotics Research, 1999
- Predicting error in rigid-body point-based registrationIEEE Transactions on Medical Imaging, 1998