Magnetization structure contrast based on intermolecular multiple‐quantum coherences
- 3 December 2002
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 48 (6) , 973-979
Abstract
In vivo and ex vivo MRI based on intermolecular multiple‐quantum coherences (iMQC) is predicted to provide a fundamentally different source of contrast for MRI. This article investigates the dependence of image contrast upon the choice of correlation distance for a heterogeneous material. A closely packed array of parallel hollow cylinders was used to demonstrate signal intensity variations when the correlation distance becomes comparable to the gap size between the cylinders. The observed effects agree well with three‐dimensional calculations of the time evolution of magnetization under the nonlinear Bloch equations. Magn Reson Med 48:973–979, 2002.Keywords
This publication has 20 references indexed in Scilit:
- Imaging the Long-Range Dipolar Field in Structured Liquid State SamplesJournal of Magnetic Resonance, 2001
- Intermolecular zero-quantum coherence imaging of the human brainMagnetic Resonance in Medicine, 2000
- Equivalence between the “classical” and the “Warren” approaches for the effects of long range dipolar couplings in liquid nuclear magnetic resonanceThe Journal of Chemical Physics, 2000
- MR Imaging Contrast Enhancement Based on Intermolecular Zero Quantum CoherencesScience, 1998
- Quantum treatment of the effects of dipole–dipole interactions in liquid nuclear magnetic resonanceThe Journal of Chemical Physics, 1996
- Structural Investigations with the Dipolar Demagnetizing Field in Solution NMRPhysical Review Letters, 1996
- Homogeneous NMR Spectra in Inhomogeneous FieldsScience, 1996
- Demagnetization field effects in two-dimensional solution NMRConcepts in Magnetic Resonance, 1996
- Intermolecular multiple-quantum coherences and cross correlations in solution nuclear magnetic resonanceThe Journal of Chemical Physics, 1993
- NMR multiple echoes observed in solidPhysical Review B, 1979