Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine Heart
Authors:
- 1 December 2014
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 7 (6) , 1174-1180
Abstract
Background—: Electrode-tissue contact force (CF) is believed to be a major factor in radiofrequency lesion size. The purpose of this study was to determine, in the beating canine heart, the relationship between CF and radiofrequency lesion size and the accuracy of predicting CF and lesion size by measuring electrogram amplitude, impedance, and electrode temperature. Methods and Results—: Eight dogs were studied closed chest. Using a 7F catheter with a 3.5 mm irrigated electrode and CF sensor (TactiCath, St. Jude Medical), radiofrequency applications were delivered to 3 separate sites in the right ventricle (30 W, 60 seconds, 17 mL/min irrigation) and 3 sites in the left ventricle (40 W, 60 seconds, 30 mL/min irrigation) at (1) low CF (median 8 g); (2) moderate CF (median 21 g); and (3) high CF (median 60 g). Dogs were euthanized and lesion size was measured. At constant radiofrequency and time, lesion size increased significantly with increasing CF ( P <0.01). The incidence of a steam pop increased with both increasing CF and higher power. Peak electrode temperature correlated poorly with lesion size. The decrease in impedance during the radiofrequency application correlated well with lesion size for lesions in the left ventricle but less well for lesions in the right ventricle. There was a poor relationship between CF and the amplitude of the bipolar or unipolar ventricular electrogram, unipolar injury current, and impedance. Conclusions—: Radiofrequencylesion size and the incidence of steam pop increase strikingly with increasing CF. Electrogram parameters and initial impedance are poor predictors of CF for radiofrequency ablation.Keywords
This publication has 14 references indexed in Scilit:
- A novel radiofrequency ablation catheter using contact force sensing: Toccata studyHeart Rhythm, 2012
- Area Under the Real‐Time Contact Force Curve (Force–Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile ModelJournal of Cardiovascular Electrophysiology, 2010
- Relationship Between Catheter Forces, Lesion Characteristics, “Popping,” and Char Formation: Experience with Robotic Navigation SystemJournal of Cardiovascular Electrophysiology, 2009
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- RF Catheter Ablation: Lessons on LesionsPacing and Clinical Electrophysiology, 2006
- Electrode Impedance: An Indicator of Electrode-Tissue Contact and Lesion Dimensions During Linear AblationJournal of Interventional Cardiac Electrophysiology, 2000
- The LETR‐Principle: A Novel Method to Assess Electrode‐Tissue Contact in Radiofrequency AblationJournal of Cardiovascular Electrophysiology, 1998
- The Effects of Electrode‐Tissue Contact on Radiofrequency Lesion GenerationPacing and Clinical Electrophysiology, 1997
- Relation between impedance and endocardial contact during radiofrequency catheter ablationAmerican Heart Journal, 1994
- Determinants of Lesion Size During Radiofrequency Catheter Ablation: The Role of Electrode‐Tissue Contact Pressure and Duration of Energy DeliveryJournal of Cardiovascular Electrophysiology, 1991