Single electron pumps: towards a quantum representation of the ampere

Authors:
  • SGS. P. Giblin
    S. P. Giblin
  • MKM. Kataoka
    M. Kataoka
  • JFJ. D. Fletcher
    J. D. Fletcher
  • PSP. See
    P. See
  • TJT. J. B. M. Janssen
    T. J. B. M. Janssen
  • JGJ. P. Griffiths
    J. P. Griffiths
  • GJG. A. C. Jones
    G. A. C. Jones
  • IFI. Farrer
    I. Farrer
  • DRD. A. Ritchie
    D. A. Ritchie

Abstract
Electron pumps generate a macroscopic electric current by controlled manipulation of single electrons. Despite intensive research towards a quantum current standard over the last 25 years, making a fast and accurate quantised electron pump has proved extremely difficult. In this article we demonstrate that the accuracy of a semiconductor quantum dot pump can be dramatically improved by using specially designed gate drive waveforms. Our pump can generate a current of up to 150 pA, corresponding to almost a billion electrons per second, with an experimentally demonstrated current accuracy better than 1.2 parts per million (ppm) and strong evidence, based on fitting data to a model, that the true accuracy is approaching 0.01 ppm. This type of pump is a promising candidate for further development as a realisation of the SI base unit ampere, following a re-definition of the ampere in terms of a fixed value of the elementary charge.