Mjölnir: a miniature triaxial rock deformation apparatus for 4D synchrotron X-ray microtomography
Open Access
- 16 October 2020
- journal article
- research article
- Published by International Union of Crystallography (IUCr) in Journal of Synchrotron Radiation
- Vol. 27 (6) , 1681-1687
Abstract
An X-ray transparent experimental triaxial rock deformation apparatus, here named `Mjölnir', enables investigations of brittle-style rock deformation and failure, as well as coupled thermal, chemical and mechanical processes relevant to a range of Earth subsurface environments. Designed to operate with cylindrical samples up to 3.2 mm outside-diameter and up to 10 mm length, Mjölnir can attain up to 50 MPa confining pressure and in excess of 600 MPa axial load. The addition of heaters extends the experimental range to temperatures up to 140°C. Deployment of Mjolnir on synchrotron beamlines indicates that full 3D datasets may be acquired in a few seconds to a few minutes, meaning full 4D investigations of deformation processes can be undertaken. Mjölnir is constructed from readily available materials and components and complete technical drawings are included in the supporting information.Keywords
Funding Information
- Natural Environment Research Council (NE/R001693/1)
- Advanced Photon Source, US Department of Energy (DE-AC02-06CH11357)
This publication has 9 references indexed in Scilit:
- Catastrophic Failure: How and When? Insights From 4‐D In Situ X‐ray MicrotomographyJournal of Geophysical Research: Solid Earth, 2020
- Rupture Energetics in Crustal Rock From Laboratory‐Scale Seismic TomographyGeophysical Research Letters, 2019
- A new mini-triaxial cell for combined high-pressure and high-temperature in situ synchrotron X-ray microtomography experiments up to 400°C and 24 MPaJournal of Synchrotron Radiation, 2019
- An experimental platform for triaxial high-pressure/high-temperature testing of rocks using computed tomographyReview of Scientific Instruments, 2018
- A deformation rig for synchrotron microtomography studies of geomaterials under conditions down to 10 km depth in the EarthJournal of Synchrotron Radiation, 2016
- Seismic attenuation in partially saturated rocks: Recent advances and future directionsThe Leading Edge, 2014
- A low-cost X-ray-transparent experimental cell for synchrotron-based X-ray microtomography studies under geological reservoir conditionsJournal of Synchrotron Radiation, 2013
- X-ray microtomography for studying localized deformation in fine-grained geomaterials under triaxial compressionComptes Rendus Mécanique, 2004
- Quasi-static fault growth and shear fracture energy in graniteNature, 1991